9103 Picoammeter bias modes

9103 Picoammeter bias modes

This post explains the 9103 picoammeter bias modes and also explains the concept of “floating”.

There are 4 picoammeter bias modes:

No bias – the current source is connected directly to the picoammeter input

No_bias_functionality

Internal – a low noise DC supply (two 45 V DC batteries in series) provides a bias between the current source and the input of the picoammeter. This application is typically used for providing a +90V bias to a target in a vacuum chamber in order to prevent secondary electrons from leaving the sample. Adding the 90V bias results in a more accurate measurement for both electron and ion current. It is also possible to use regular 9V batteries and so lower internal bias voltages are possible. Batteries are used for the internal bias as they are very clean with no ripple.

Internal_bias_functionality

 

 

 

 

 

External –

There are two external bias options for the 9103, both are limited to 600 V DC.

9103EXTBIASPR has a BNC connector on the back of the 9103 that connects to the customer’s voltage source which needs to be isolated from ground (floating).

9103DBPR (shown below) has two BNC connectors on the front panel of the 9103 and the 9103 floats on the customer’s ground referenced DC voltage source. 

The 9103DBPR is the most common external bias option as most voltage sources are referenced to ground (not isolated).

Two BNC connectors are used to float the 9103 up to as much as +/- 600V DC with an external low noise voltage source (provided be the user).    The voltage source is connected to the HV (signal ground) BNC connector and the 9103 and the current source (DUT) is connected to the Input BNC connector.

The maximum voltage that can be applied in this mode is 600V DC (that is the rating of the BNC input on the 9103 USB picoammeter).

external bias

External Bias Dual BNC

Floating – In this mode the picoammeter signal ground is raised from near chassis ground up to a value as high as +/- 5000V DC.   If you were to measure between the signal input and the signal ground reference the voltage would be low (under 1V typically). However measuring the DC voltage from the signal reference to chassis ground the voltage would be high, the floating supply value.

When floating the 9103 USB picoammeter there are some safety considerations. First of all, the input to the 9103 is changed from a BNC to either a SHV or MHV connector (user specified). These connectors are rated for up to 5kV. The recorder output on the back of the 9103 also floats. For example if you apply a floating voltage of 1000V the recorder output would also measure 1000V with respect to chassis ground. So if you are using the recorder output and also floating the 9103 the meter that you are measuring the recorder output with also needs to be isolated from ground. Finally, to prevent voltage spikes a current limit resistor is placed between the floating supply and the 9103 signal ground.   The floating mode is typically used to measure the collector output of an electron multiplier or a Faraday cup.

The maximum float voltage with the latest rev 9103 is +/- 5000 V DC.   If your 9103 was purchased before January 2019 then it can only float up to 1500V DC maximum.

The floating mode cannot be used with either the internal or external bias modes.  However you really would not need either of those options as you can float the 9103 up to get what ever bias you need.  For example,  let’s say you want to measure an ion beam that is at 3000V DC and you wanted to retard the ions by 200V.  You could simple set the bias supply to +3200 V DC and you would  be retarding the 3000 V beam by 200V DC.

For all modes of operation the chassis ground of the 9103 USB picoammeter should be connected to the ground on the PC or chamber.

If you would like more detailed information please contact us via email at the link in the upper right hand corner of our website at www rbdinstruments dot com.

9103 Actuel Beta: Data Cursor and Auto-save

UPDATE: A new Actuel Beta featuring a data cursor and auto-save is now available.

Actuel, RBD’s Windows application for controlling the 9103 USB Picoammeter, has currently been updated with some great new features. While the official release is a few weeks away, the beta release contains some new features in addition to the previously previewed micro-window display.

Auto-Save while recording:

In response to customer requests, we’ve added a feature to automatically save data while recording. The auto-save feature comes in handy when leaving the 9103 unattended for long periods. While rare, brownouts, etc., can cause a temporary communication failure between the 9103 and PC, and USB failures can sometimes lock-up or crash a system. Auto-save ensures that data is stored to a text file every 5 minutes, and again when recording is stopped (for completeness).

Auto-save always works automatically when multiple 9103 picoammeters are synced together. Simply select the option for all instances of Actuel, then select the file name (and keep the Data window open for each instance you want to record/save from). The data from each 9103 will be appended to the file sequentially, separated by the device identifier.

9103 Actuel Auto-save Feature

Data Cursor:

Sometimes it’s handy to see the exact data at a particular point in time during or after data acquisition. While high sampling rates make it difficult to mouse-over every data point in large data sets, the data cursor can help track current vs. time more easily without the need to export data as text. By simply hovering over the graph, a vertical line appears, while current and time is displayed in the upper right-hand corner of the graph window.

actuel-data-cursor

As always, Actuel is free for all owners of a 9103 Picoammeter. Meanwhile, you can download the most recent version of Actuel here:

Download Actuel for the 9103 USB Picoammeter

More Information on the 9103 USB Picoammeter

Calibrating the 9103 USB Picoammeter

9103 Picoammeter

Calibrating the 9103 USB Picoammeter

The calibration procedure provided with the 9103 USB Picoammeter is written for use with a Keithley 220 current source and is set up for semi-automatic operation. But what if you want to use another brand or type of current source? The following procedure shows you exactly how to calibrate and verify calibration using any current source.

Preparing for Calibration

In order to calibrate your 9103 Picoammeter, you will need a calibrated current source, and software that can send / receive ASCII commands to the Picoammeter. If you are using the Actuel application supplied with your Picoammeter, there is a built in console window that allows you to communicate directly with the Picoammeter using ASCII commands – simply click the Console button to open the window and type the commands in the field provided.

Note that when using Actuel’s console window, the ampersand character “&” that precedes all commands is automatically entered for you – you do not have to type it.

Note also that the offset and gain calibration apply to all 7 ranges. Range 1 corresponds to the highest resolution range in nA, range 7 corresponds to the lowest resolution range in mA. The Gain Reference Table (at the end of this article) shows the ideal calibration values and limits for each range (both positive and negative polarity).

Perform OFFSET Calibration and Verification Calibrate Offset (all ranges)

  1. Ensure that the current input is disconnected
  2. Send command &R0 (auto-range)
  3. Send command &C1, 9103 responds “Offset Calibration starting…”
  4. Wait for completion (about 30 seconds), until message received “Offset Calibration Completed!”
  5. Send command &C0, 9103 responds “OPEN CIRCUIT Offset Calibration starting…”
  6. Wait for completion (about 30 seconds), until message received “OPEN CIRCUIT Offset Calibration Completed!”

Verify Offset for (all ranges)

  1. Ensure that the current input is disconnected
  2. Send the &I0500 command to start receiving data samples
  3. Send command &R# (where # is a number 1 through 7) to manually select the range to be verified (see the Gain Reference Table below)
  4. Verify measured values received are zero and have at least 1 zero after the decimal point
  5. Repeat until all ranges are checked

Perform GAIN calibration

Refer to the Gain Reference Table below for interpretation of the current values and limits.

  1. Send the &+CC command to unlock the calibration function. The green LED will begin flashing
  2. Connect the current source to the 9103 input
  3. Turn on current source. Ensure current source is at proper operating temperature before proceeding
  4. Send command &C2, and wait for prompt from 9103: “Apply test current…”
  5. Using your calibrated current source, input the specified test current (see table below) and send command &C2 again, the 9103 will respond with “Measuring and calibrating gain…”
  6. Repeat the current application for each of the 7 ranges and polarities, following the prompts provided by the 9103
  7. Wait for response from the 9103: “Gain Calibration completed!”

Perform GAIN verification

  1. Send command &R0 (select auto-ranging)
  2. Send command &I0500 (sample interval 500 mSec)
  3. For each test current (shown in the Gain Reference Table below), set the source to the specified current and select the appropriate current range on the 9103 in turn (&R1, &R2, &R3, &R4, &R5, &R6, &R7), monitoring the data stream for at least 10 seconds after each range, and ensuring that the readings are stable in each range (if a message starting with &S* appears indicating an unstable value, wait a little longer)
  4. Verify that the measured value reported by the 9103 is within 1% of the set current for each range
  5. Send command &I000 to stop sampling
  6. Disconnect current source and turn off the 9103 – calibration is complete

Gain Reference Table
(14 test currents are shown, positive/negative polarities for each of the 7 ranges)

Range

Test Current

9103 Value

9103 Value Limits

1

+ 1.0000 –9

1 nA

0.9900, 1.0100 nA

2

+ 10.000 –9

10 nA

09.900, 10.100 nA

3

+ 100.00 –9

100 nA

099.00, 101.00 nA

4

+ 1.0000 –6

1 uA

0.9900, 1.0100 uA

5

+ 10.000 –6

10 uA

09.900, 10.100 uA

6

+ 100.00 –6

100 uA

099.00, 101.00 uA

7

+ 1.0000 –3

1 mA

0.9900, 1.0100 mA

1

1.0000 –9

1 nA

-0.9900, -1.0100 nA

2

10.000 –9

10 nA

-09.900, -10.100 nA

3

100.00 –9

100 nA

-099.00, -101.00 nA

4

1.0000 –6

1 uA

-0.9900, -1.0100 uA

5

10.000 –6

10 uA

-09.900, -10.100 uA

6

100.00 –6

100 uA

-099.00, -101.00 uA

7

1.0000 –3

1 mA

-0.9900, -1.0100 mA