32-100 Repair & Calibration

32-100 Repair & Calibration

The 32-100 electron multiplier supply is used on many older PHI Auger and XPS surface analysis systems. This post has information about the 32-100 repair and calibration.

Command Structure

Use the diagnostic dialog box in Augerscan software to load the digital values per the 32-100 calibration procedure when calibrating the 32-100 in the digital mode.  CAUTION! Note that only personnel qualified to work safely with high voltage should perform these calibrations and tests.  Voltages of up to 5,000V DC are possible in the 32-100 electron multiplier supply.

Function                                               Command/Data(hex)

SEDVoltage                                       Addd

CMAVoltage                                       Cddd

Notes:   – the command and the data are combined in a 16 bit word

–  the left-most digit is most significant in address position and data value

–  “d” represents a data value in hex

– 000H = 0, 7FFH = 2047 (mid-scale), FFFH = 4095(max.)

Clear the CSR (= 0) for proper communication.

 Initial Setup

 Install a jumper wire on the back panel from IC (J7) to ground (J6). Note: if you do not do this, the CMA voltage will float (not referenced to chassis ground). Set the CMA and SED high voltage controls on the front panel completely counterclockwise and set the mode switches to “Off”. Disconnect P7 and P8 on the board.

DIP Switches:

  • SW1: 1 and 4 closed
  • SW2: 5 closed
  • SW3: 1 closed
  • all others open

Calibration

 Low Voltage  Power Supplies

Turn on the power. Do the following checks and adjustments with the reference to chassis ground. Always check for significant AC ripple on any DC power supplies.

Check for :

  • +15Vdc onJ7-5
  • -15Vdc onJ7-4
  • +5Vdc onJ8-1

Make the following adjustments, checking for any significant AC ripple:

  • Adjust R83 for 10.00Vdc on TP2.
  • Adjust R79 for +17Vdc on P6-2.
  • Adjust R9 for +300Vdc at E4.
  • Adjust R22 for +300Vdc at E2. (Note, adjust this to 50V if you have a 20-805 or 20-810 control on an double pass XPS system)

Note that for the above measurements, the chassis is the ground reference.

SED HV Supply

Using a HV probe, measure the voltage at E3 for the following conditions. The reference is to chassis ground. Note: the bit number in () indicates the only bit set (=1), for troubleshooting purposes.

Start by calibrating the HV limit (AFFFH=3300V). Adjust R54 to calibrate. Verify that 3300V is also achieved by turning the SED HV control knob completely clockwise with the mode set to“Analog”.

 

Command/Data E3 voltage (V)
A000H 430
A100H     (bit8) 670
A200H     (bit9) 860
A400H     (bit10) 1220
A800H     (bit11) 1920
AFFFH 3300

 

Check for increasing voltage as you enter the following values in sequence: A000H, A00FH (bits 0-3), A010H (bit 4), A020H (bit 5), A040H (bit 6), A080H (bit 7).

CMA HV Supply

Using a HV probe, measure the voltage at E1 for the following conditions. The reference is to chassis ground. Remember to connect the jumper from J7 to J6 on the back panel.

Start by calibrating the HV limit (CFFFH=4000V). Adjust R62 to calibrate. Verify that 4000V is also achieved by turning the CMA HV control knob completely clockwise with the mode set to“Analog”.

Note: the bit number in () indicates the only bit set (=1), for troubleshooting purposes.

 

Command/Data E1 voltage (V)
C000H 430
C100H     (bit8) 720
C200H     (bit9) 950
C400H     (bit10) 1400
C800H     (bit11) 2280
CFFFH 4000

Check for increasing voltage as you enter the following values in sequence: C000H, C00FH (bits 0-3), C010H (bit 4), C020H (bit 5), C040H (bit 6), C080H (bit7).

Troubleshooting Tips

  1. One of the first things to check is the +17Vdc power supply to the two TIP120 transistors when the 32-100 power is switched on. If it drops significantly (typically down to just a few volts), then the problem is likely one of the HV transformers (T1 or T2) or the clock. Disconnect jumpers J29 and J30 (short wires next to the TIP120s). Turn on the 32-100 power and check for clock pulses at TP8, TP9, TP10, and TP11. If the clock is not detectable at any or all of these test points, then there is a problem in the clock circuitry. If the clock is okay, then test each transformer one at a time by connecting one of the jumpers, turning on the 32-100 power, and noting if the +17Vdc supply drops. Replace the transformer if it causes the +17Vdc supply to drop when its jumper is connected (J29 for the SED and J30 for the CMA).
  2. Another common problem involves the low voltage circuitry for each supply (drivers for E2 and E4). If these points are not at 300Vdc and cannot be adjusted, then the problem lies here. The most likely components to fail are the diodes, especially the 1N970s (CR19 and CR27).
  3. Commonly needed repair parts:

Transistors: TIP50, TIP120, MPF108, 2N3904

Diodes: 1N970A, 1N972A,1N459A

CAUTION! Note that only personnel qualified to work safely with high voltage should perform these calibrations and tests.  Voltages of up to 5,000V DC are possible in the 32-100 electron multiplier supply.

If you are unable to perform these tests RBD Instruments can repair your 32-100 electron multiplier supply and also provide a loaner unit if needed.

32-100 electron multiplier supply digital mode

The model 32-100 electron multiplier supply is used on older Physical Electronics Auger electron spectroscopy and X-ray photoelectron spectroscopy surface analysis systems to control the electron multiplier voltage.

When using the 32-100 electron multiplier supply in the digital mode (the software automatically sets the electron multiplier voltage)  the auto-ems box in the Auger Scan software AES electron multiplier properties dialog box needs to be checked, and the CMA multiplier switch on the front panel of the 32-100 must be set to digital.

But what if the 32-100 still does not work in the digital mode?  In that case, there may be a problem with one of the digital ICs.  The following procedure explains how to try and repair this problem.

  1. Turn off the 32-100 and remove the cover.
  2. Most 32-100s have three ICs for the SED digital side of the control which are not used. You can move those ICs over to the CMA side of the control and see if that solves the problem.
  3. Remove ICs U2, U4 and U6. These are the CMA digital ICs. Then, move over U3, U5 and U7 from the SED side if those chips are available. If they are missing, then you will need to order some of those ICs from Digikey, Newark or RBD Instruments. U2 and U4 are 74LS174s and U6 is an AD7521.
  4. If you did have those ICs and that did not solve the problem, then it most likely means that one of the encoder circuit ICs are defective. Those are U16 (74123) and U1 (AM25LS2538). RBD Instruments provides these parts.
  5. You do not need to replace the SED digital ICs as they are not used.

It is assumed that the 32-100 is working in the analog mode. If the 32-100 is not working in the analog mode it will not work in the digital mode either.

Refer to the pictures below for the locations of the ICs on the 32-100 motherboard.

Note that the switch positions need to be set as shown below.

 

32100-switch-positions

32100-switch-positions

 

If you need further help troubleshooting your 32-100 electron multiplier supply please contact RBD Instruments dot com

This slideshow requires JavaScript.