20-805 or 11-500A analyzer control sweep voltage test
Older PHI cylindrical mirror analyzers (CMA) uses two cylinders to focus electrons of a particular energy into an aperture for amplification via an electron multiplier.
For AES, the inner cylinder is at ground potential and the outer cylinder has a voltage applied to it (sweep voltage) that deflects the electrons into the collection aperture .
This blog post will show you how to test the sweep voltage.
11-500A Analyzer Controller
Turn off the 11-500A analyzer controller.
Remove the VM cable from the analyzer.
Set up an alignment with the lower limit at 100eV and the upper limit also set to 100eV. This will set the sweep voltage to sit at 100eV.
Remove the VM cable from the analyzer.
Connect your DVM meter to the center pin on the end of the VM cable with respect to the vacuum chamber.
Turn on the 11-500A and measure the DC voltage on the end of the VM cable.
Acquire the alignment. The scale factor for the analyzer is 1.7 so the sweep voltage for 100eV is 100 divided by equals 58.823 VDC. If you have close to 58.8V DC on the end of the VM cable, then the 11-500A sweep voltage is working properly.
20-805 Analyzer Controller
15-255G Analyzer
Turn off the 20-805 analyzer controller. Make sure that it is in the AES mode.
Remove the OC cable from the analyzer.
Set up an alignment with the lower limit at 100eV and the upper limit also set to 100eV.
Turn on the 20-805 and acquire the alignment. Measure the DC voltage on the end of the OC cable with respect to the vacuum chamber. You should have about 58.5V DC on the end of the OC cable. If so, then the sweep voltage is working properly.
25-260/270 Analyzer
Turn off the 20-805 analyzer controller. Make sure that it is in the AES mode.
Remove the Analyzer cable from the CMA.
Pin B on the analyzer connector is the OC voltage.
Set up an alignment with the lower limit at 100eV and the upper limit also set to 100eV.
Turn on the 20-805 and acquire the alignment. Measure the DC voltage on pin B of the OC cable with respect to the vacuum chamber. You should have about 58.5V DC on the end of the OC cable. If so, then the sweep voltage is working properly
If the sweep voltage is not working then contact RBD Instruments for assistance.
16-050 Heat Exchanger Flow Switch replacement procedure.
The original flow switch for the 16-050 heat exchanger used 220 VAC for the input power.
About 10 years ago that flow switch was discontinued and was replaced with a similar model that used 120 VAC for the input power. A 220 VAC to 120 VAC step down transformer was also needed.
Very recently, the 120 VAC input power flow switch was discontinued, and it has been replaced by another version that uses 24 VDC as the input power. To provide 24V DC to the replacement flow switch a small 220VAC to 24 V DC power supply is needed.
This blog post will provide step by step instructions on how to install the new 24V DC flow switch and power supply into a 16-050 heat exchanger.
Step one is to drain the water from the heat exchanger. The easiest way to do that is to remove the IN and OUT couplers from the X-ray source and then connect those to the IN and OUT lines. Mark the IN and OUT water lines to make sure that you do not get them mixed up. Hold them over a clean container (such as a plastic waste basket) and then turn on the 32-095 or 32-096 X-ray source control.
The 16-050 will turn on and water will come out of the source IN line. After about 9 seconds the 32-095/96 will shut down the 16-050 heat exchanger due to insufficient flow. Turn the 32-095/96 OFF for a few seconds, then back ON. Repeat several times (it might take 10 to 15 times) until water stops coming out of the source IN line and the pump starts to make noise. This will drain most of the water out of the 16-050. You want to use a clean container for the water so that you can reuse the water. Or you can replace the water after the flow switch has been replaced with distilled water from a grocery store. Lab DI water does not always work.
After the water has been drained, turn the 16-050 heat exchanger front panel power switch to OFF.
Next, trace the power cord that is on the back of the 16-050 into the electronics console and unplug the power cord. Make a note of where that power cord plugs into so that you can plug it into the same receptacle later.
Turn the 16-050 front panel power switch to ON just to make sure that the 16-050 power cord is unplugged. Verify that the 16-050 motor does not turn on and then turn the 16-050 heat exchanger front panel power switch back to OFF. Do not work on the 16-050 if it is still plugged in!
Remove the top cover on the 16-050.
Remove the small cover that is located just behind the 16-050 power switch. There are 4 nuts that need to be removed then the cover comes off.
Remove the wires that go to the flow switch. Those will be TB1 sections 1, 2, 5 and 7.
Remove the old flow switch. Use your phone to take pictures of where the IN and OUT lines connect on the flow switch.
Install the new flow switch making sure that it is connected EXACTLY the way that the old flow switch was.
Route the new flow switch wires up to where the old flow switch wires were.
The wires need to be connected as follows:
RAVCV04-24 power supply:
Blue to TB1 7
Brown to TB1 5
Red to Flow Switch Brown
Black to Flow Switch White
Flow Switch:
White to RACV04-24 Black (use twist nut)
Brown to RACV04-24 Red (use twist nut)
Red to TB1 2
Black to TB1 1
Green – Not used.
This drawing shows the connections –
Verify that the wires are all tight and then replace the little cover using the 4 nuts.
Reconnect the couplers and water lines to the X-ray source. Make sure that you have the IN and Out going to the correct ports on the X-ray source.
Pour most of the water back into the reservoir or use 2-3 gallons of distilled water from a grocery store.
Plug the 16-050 power cord back in and then turn on the 16-050 front panel power switch.
Turn on the 32-095/96 and the 16-050 motor should turn on. It will turn off after 9 seconds. Turn the 32-095/06 off and then back on to reset the pump. Do this a few times until the air bubbles work themselves out and the 16-050 pump motor stays on.
You will need to top off the water reservoir once the water fills up the lines. Do not overfill!
If the 16-050 will not stay on even when the water is full (close to the upper line on the water level indicator on the back of the 16-050) then the flow switch may need to be adjusted.
The flow switch adjustment potentiometer is on the back of the flow switch under a label. It is a small, slotted screw.
Try turning the flow switch a few turns CW and then turn on the 32-095/96. If that does not work, try a few turns CCW plus a few more turns. Once you get it to where the 16-050 pump stays ON then you can slowly turn the potentiometer the opposite direction until the red interlock light on the 32-095/96 turns ON, then adjust the potentiometer 1.5 turns the other direction until the red interlock light just turns off. The idea is that you want to set the potentiometer so that the 16-050 pump is ON when you have water flow, but you want to be somewhat close to the edge so that the 16-050 pump will shut off when the flow drops.
Typical flow rates are 1.1 GPM for XPS systems with a 10-610 mono-chromator source and up to 1.8 GM for systems that have only a single 04-548 dual anode X-ray source.
The RBD Instruments Part number for the kit that includes the 24V flow switch and the power supply is FLOW-SWITCH-24-Kit.
For more information please contact RBD Instruments here – https://rbdinstruments.com/contact.html