Protect the CF flange knife edge

The CF flange (also commonly referred to as the Varian trade-marked Conflat) uses a knife edge that cuts into a soft metal gasket (usually copper) to provide a leak-tight metal-to-metal seal. “Protect the knife edge at all costs” is a common catchphrase at RBD Instruments that we like to drill into anyone who works on a UHV vacuum chamber. Simply put, it means to use care when removing used copper gaskets from Conflat (CF) flanges, and to protect the flange edge with aluminum foil when it is being stored on a shelf. Even a small scratch on the knife edge can result in a leaky seal on the flange.

Specialized tools are available for removing copper gaskets, but most labs do not have them. So, what is the best way to remove a copper gasket from a CF flange if you do not have a special tool? IMO you should use a clean regular channel lock or vice grip type of pliers whenever possible to remove used copper gaskets instead of a flat blade screwdriver. You need to use pliers that have teeth, not the smooth ones. In some cases you will not be able to use pliers as there is not enough room to get the pliers onto the gasket without risking damage to a ceramic or some other optic part. In those cases, you will need to use a straight blade screwdriver.

To use pliers, you simply position the jaws of the pliers on the inner and outer diameters of the gasket and tilt the pliers back. The applied leverage will pop the gasket out of the groove. Be sure to get a good grip on the edge of the gasket as quite often the gasket is pressed in very tightly and will be somewhat difficult to remove. Other times the gaskets will practically fall out by themselves. It just depends on the fit of the flanges and whether or not the system has been baked out (the copper expands during bake-out).

When using a screwdriver, extreme care must be taken to ensure that the screwdriver does not slip off the side of the gasket and damage the knife edge once the gasket pops out of the flange recess. You can reduce the chances of damage by using a screwdriver that has a tip that is flat with sharp edges and slightly thinner than the height of the exposed gasket edge. Apply downward pressure on the tip of the screwdriver as you try to pry the gasket up. Sometimes the gaskets can be wedged in very tightly after compression and be difficult to remove. Use the tip of the screwdriver to leverage the gasket out; do not push the screwdriver. This is when one is most likely to damage a knife edge: if the screwdriver slips as the gasket comes loose, it may scrape across the knife edge and cause a large scratch or gouge. Once that happens, the flange may no longer seal properly.

If you do gouge a knife edge you might be able to still get it to seal by using a small piece of a soft vacuum-compatible metal such as gold or platinum to fill in the void on the gouge. Small scratches can be polished out with some ultra-fine emery cloth. If the gouge is too big then the flange may need to have the knife edge touched-up with a metal lathe. That is not always practical or even possible. So, the best thing to do is to… protect the knife edge at all costs.

When flanges are being stored for long periods of time or are being stacked up on a shelf, the best way to protect them is to wrap them in UHV aluminum foil. You can purchase UHV compatible aluminum foil from All-Foils Inc. http://www.allfoils.com/single-product/uhv-foil/

Below are some pictures of the correct and incorrect techniques to remove copper gaskets from CF flanges. If you need to purchase some copper gaskets for your UHV vacuum chamber and also want to pay the lowest possible price, RBD Instruments Inc. now provides copper gaskets for CF flanges in small or large quantities at low prices.  Click on this link for more information // Copper gaskets for CF flanges // or visit us at our website by going to RBD Instruments dot com.

Ion Plasma to clean ion pumps

If you try to start an ion pump when the vacuum in the chamber is in the mid 10-4 range, the gas load will be high enough to produce a visible ion plasma.  Normally you don’t start the ion pumps until the vacuum is pulled down to the low 10-5 range by the turbo pump.  But, sometimes you want to deliberately generate an ion plasma to help clean the ion pump elements.

http://youtu.be/vqTTybwSDl0

There are two ways to do this.

  1. Just start the ion pumps when the vacuum reaches the mid to low 10-4 Torr range.  You may see that the pressure in the chamber rises to the 10-3 Torr range when the ion pump high voltage is turned on. That is OK; keep the ion pumps on while pumping the chamber with the turbo pump.  You can leave them on for 5 minutes or so, then shut off the ion pump supply and let them cool down for 5 minutes. Then repeat the process. After a number of cycles, vacuum will be in the low 10-5 range and the ion pumps will start. You know when the ion pumps start because the vacuum goes into the 10-6 range and keeps improving slowly. By forcing the ion pumps to start in the high 10-4 range the resultant ion plasma helps to clean the ion pump elements.
  2. If the pumps are loaded with argon or contaminated with hydrocarbons, you want to use oxygen to produce the ion plasma because oxygen will react with the contaminants.  Assuming the ion pumps are started, back fill oxygen into the vacuum chamber to 5 X 10-5 Torr.  Turn off the ion gauge and monitor the current on the ion pump control.  Increase the oxygen until you get about 50mA of current on the ion pump control.  Adjust the oxygen leak valve as needed to maintain 50mA or so of current.  Maintain this condition for about 30 minutes, and then turn the oxygen off.  As the pumps cool down the vacuum will recover and typically by the next day the ion pumps are happy once again.

For more info on ion pumps type Ion Pump Element rebuild procedure in the RBD TechSpot search box

Varian variable leak valve adjustment procedure

This post is about how to adjust the 951-5100 and 951-5106 Varian variable leak valve commonly found on vacuum chambers across the globe.  Invented in 1968 by William Wheeler and Paul Hait, the patent was assigned to Varian Associates; many thousands of these valves have been built since then. In 2010, Varian was purchased by Agilent Technologies and since that time it has been harder to find the rebuilt kits for these variable leak valves. The order information for the rebuild kit can be found at the bottom of this post, as well as a link to the instruction manual.

Varian variable leak valve

Varian variable leak valve

Overview-

When this type of variable leak valve does not close all the way, operators will usually unlock the two knurled knobs and move them up the fine drive screw assembly and then re-lock them.  What that does is to allow the handle to move out further from the valve body and put more pressure on the sapphire, which effectively closes the valve.  That works fine for periodic adjustments to close the valve. However, if the handle is far away from the valve body it is possible to get too much leverage and crack the sapphire (which causes catastrophic valve failure) and also reduces the range of the valve arm. The picture below shows a valve handle that is too far away from the valve body and needs to be reset.

misadjusted varian leak valve

Adjustment procedure –

When properly adjusted, the handle will be parallel to the valve body when closed. Gas should just start to leak into the vacuum chamber when the knob is turned CCW by 2 full turns.  The following procedure is an abbreviated version that I use which works most of the time. For a more thorough explanation of this adjustment refer to the manual.

CAUTION!   Read and understand the procedure and notes before you attempt the adjustment on your system.  You need to know exactly how much gas pressure you are putting into the valve, how much you need to adjust the valve by and the status of your vacuum chamber.  The roughing screw is very sensitive; a very small amount of adjustment can be the difference between success and a dumped system!

1. Use a small straight blade screwdriver to remove the hole cover on the back end of the leak valve to expose the roughing screw cap hex head.
use screwdriver to remove valve endcap

2. Insert a 5/16” Allen wrench into the cap head of the roughing screw

allen wrench 5-16

use hex head allen wrench to adjust roughing screw

3. Monitor the vacuum in the chamber. For most PHI surface analysis systems the vacuum should be in the low 10-9 Torr or lower.
4. If necessary, adjust the handle so that it is parallel to the valve body. The knobs will need to be adjusted – turn the top knob CCW while holding the bottom knob in place. That will separate the two knobs and then you can spin them one at a time into position. Lock them on the fine drive screw assembly by holding the bottom knob in place and turning the top knob CW until it firmly butts up against the bottom knob.  There is a spring washer between the two knobs that helps to lock them together.  Once set, always turn both knobs together when opening (turn CCW) or closing (turn CW) the valve.

adjust handle so that it is parallel to body assembly

NOTE: Keep an eye on the vacuum in the chamber as you SLOWLY adjust the handle. If the pressure starts to increase, turn the Allen wrench CW by a few degrees to close the valve. The roughing screw is coarse threaded and so a very small movement on the Allen wrench has a large effect on the valve, whether closing (CW) or opening (CCW). The maximum torque that should be applied to the roughing screw is 6 foot lbs.

5. Once the handle has been set so that it is parallel to the valve body and the valve is closed (no gas leaking into the chamber) you are ready to adjust the open position of the valve.
6. SLOWLY turn the knob (both knobs turn together) CCW while observing the vacuum in the chamber. You want to get to two full turns without bleeding any gas into the chamber. If you start to see leakage before two full turns on the valve, compensate by turning the Allen wrench CW slightly until the leakage stops and the vacuum starts to recover.
7. Once you have the valve open two full turns with no leakage, SLOWLY turn the Allen wrench CCW in increment of 1 to 2 degrees at a time until the gas just starts to leak into the vacuum chamber.  Then close the leak valve by turning the knob fully CW (two turns).
8. Verify that the gas just starts to leak into the vacuum chamber at about 2 turns. If necessary, adjust the Allen wrench in very small increments.  Sometimes it is not possible to have the valve open at two turns.  It may not open until as many as 6 turns, and that is still acceptable. The idea is to have predictable, smooth and repeatable control.
9. When you close the valve, go just finger tight – do not over tighten the knob or you will damage the threads on the fine drive screw! The fine drive screw should be periodically lubricated, I prefer C5A over moly-disulfide.

NOTE:  If you are adjusting the leak valve after connecting a pressurized gas bottle for the first time, have the Allen wrench inserted into the roughing screw so that you can quickly close the valve further if the pressure in the chamber rises.  Sometimes a valve will seal fine in atmosphere but leak when up to 500 PSI is applied to the back end of the valve. The valve is rated for a maximum of 500 PSI inlet gas pressure, but it will work better with 100 PSI or less.

If this procedure does not work, then the valve may need to be cleaned or the sapphire and gasket assemblies replaced. There is a limited lifetime on the gasket assembly as the copper gasket becomes compressed with each use.  The manual specifies anywhere from 20 to 300 valve closures based on whether or not the valve is baked out or not. However, on the surface analysis systems where I have seen these valves used, they can operate for many years with little or no adjustment.

Refer to the manual for cleaning and rebuild instructions as well as a more detailed adjustment procedure.

adobe_pdf Varian variable leak valve manual

Agilent (Varian) variable leak valve part numbers (refer to drawing below) and prices as of February 2013:

9620014 $546.00 Repair and tool kit includes all of the parts and tools that are required in order to completely rebuild the valve as well as instructions
9535050 $322.00 Replacement gasket assembly
9530072 $398.00 Replacement sapphire assembly
9515106 $1,372.00 The entire valve – Sapphire-sealed variable leak valve and valve adjustment toolsWith NW16 CFF gas inlet

Valve repair kit parts list_001

TO PLACE AN ORDER, Agilent offers several options:

1)      Visit http://www.chem.agilent.com/store/StoreSearchResults.aspx?sp=9620014  to place online orders using a purchase order or credit card.

2)   Call 1-800-882-7426 (option 1) any weekday between 8 am and 7 pm Eastern time in the U.S., and Canada.

3)   To place an order by telefax: 1-781-860-5405

4)   To place an order by email: [email protected]

5)   Or you can mail your order to:

Agilent Technologies, Inc

121 Hartwell Avenue
Lexington

MA 02421 USA

To place an order, the following information is required: Purchase order number or credit card, delivery date, ship to, invoice to, end user, and quote number. GSA customers please provide GSA contract #.

TO CHECK THE STATUS OF AN ORDER:

1)      Visit http://www.chem.agilent.com/en-US/Products/Instruments/vacuum/Pages/default.aspx to check the status of your order.

2)      Call 1-800-882-7426 (option 1) any weekday between 8 am and 7 pm Eastern time, in the U.S., and Canada. You will need to know the purchase order or credit card number the order was placed on.