Calibrating the 9103 USB Picoammeter

9103 Picoammeter

Calibrating the 9103 USB Picoammeter

The calibration procedure provided with the 9103 USB Picoammeter is written for use with a Keithley 220 current source and is set up for semi-automatic operation. But what if you want to use another brand or type of current source? The following procedure shows you exactly how to calibrate and verify calibration using any current source.

Preparing for Calibration

In order to calibrate your 9103 Picoammeter, you will need a calibrated current source, and software that can send / receive ASCII commands to the Picoammeter. If you are using the Actuel application supplied with your Picoammeter, there is a built in console window that allows you to communicate directly with the Picoammeter using ASCII commands – simply click the Console button to open the window and type the commands in the field provided.

Note that when using Actuel’s console window, the ampersand character “&” that precedes all commands is automatically entered for you – you do not have to type it.

Note also that the offset and gain calibration apply to all 7 ranges. Range 1 corresponds to the highest resolution range in nA, range 7 corresponds to the lowest resolution range in mA. The Gain Reference Table (at the end of this article) shows the ideal calibration values and limits for each range (both positive and negative polarity).

Perform OFFSET Calibration and Verification Calibrate Offset (all ranges)

  1. Ensure that the current input is disconnected
  2. Send command &R0 (auto-range)
  3. Send command &C1, 9103 responds “Offset Calibration starting…”
  4. Wait for completion (about 30 seconds), until message received “Offset Calibration Completed!”
  5. Send command &C0, 9103 responds “OPEN CIRCUIT Offset Calibration starting…”
  6. Wait for completion (about 30 seconds), until message received “OPEN CIRCUIT Offset Calibration Completed!”

Verify Offset for (all ranges)

  1. Ensure that the current input is disconnected
  2. Send the &I0500 command to start receiving data samples
  3. Send command &R# (where # is a number 1 through 7) to manually select the range to be verified (see the Gain Reference Table below)
  4. Verify measured values received are zero and have at least 1 zero after the decimal point
  5. Repeat until all ranges are checked

Perform GAIN calibration

Refer to the Gain Reference Table below for interpretation of the current values and limits.

  1. Send the &+CC command to unlock the calibration function. The green LED will begin flashing
  2. Connect the current source to the 9103 input
  3. Turn on current source. Ensure current source is at proper operating temperature before proceeding
  4. Send command &C2, and wait for prompt from 9103: “Apply test current…”
  5. Using your calibrated current source, input the specified test current (see table below) and send command &C2 again, the 9103 will respond with “Measuring and calibrating gain…”
  6. Repeat the current application for each of the 7 ranges and polarities, following the prompts provided by the 9103
  7. Wait for response from the 9103: “Gain Calibration completed!”

Perform GAIN verification

  1. Send command &R0 (select auto-ranging)
  2. Send command &I0500 (sample interval 500 mSec)
  3. For each test current (shown in the Gain Reference Table below), set the source to the specified current and select the appropriate current range on the 9103 in turn (&R1, &R2, &R3, &R4, &R5, &R6, &R7), monitoring the data stream for at least 10 seconds after each range, and ensuring that the readings are stable in each range (if a message starting with &S* appears indicating an unstable value, wait a little longer)
  4. Verify that the measured value reported by the 9103 is within 1% of the set current for each range
  5. Send command &I000 to stop sampling
  6. Disconnect current source and turn off the 9103 – calibration is complete

Gain Reference Table
(14 test currents are shown, positive/negative polarities for each of the 7 ranges)

Range

Test Current

9103 Value

9103 Value Limits

1

+ 1.0000 –9

1 nA

0.9900, 1.0100 nA

2

+ 10.000 –9

10 nA

09.900, 10.100 nA

3

+ 100.00 –9

100 nA

099.00, 101.00 nA

4

+ 1.0000 –6

1 uA

0.9900, 1.0100 uA

5

+ 10.000 –6

10 uA

09.900, 10.100 uA

6

+ 100.00 –6

100 uA

099.00, 101.00 uA

7

+ 1.0000 –3

1 mA

0.9900, 1.0100 mA

1

1.0000 –9

1 nA

-0.9900, -1.0100 nA

2

10.000 –9

10 nA

-09.900, -10.100 nA

3

100.00 –9

100 nA

-099.00, -101.00 nA

4

1.0000 –6

1 uA

-0.9900, -1.0100 uA

5

10.000 –6

10 uA

-09.900, -10.100 uA

6

100.00 –6

100 uA

-099.00, -101.00 uA

7

1.0000 –3

1 mA

-0.9900, -1.0100 mA

Ion source filament assembly for PHI 04-303, 06-350 and FIG 5 ion sources

RBD Instruments’ proprietary technique for rebuilding the PHI 04-303 ion gun and 06-350 ion source filament assembly results in more stable sputter rates and longer filament lifetimes than other ionizers. The pictures below are actual 04-303 ion source filament assemblies which show the visible light emitting from the filaments. Note that the 3 post RBD ionizers have a much larger and more uniform hot spot than the 5 post PHI filament. In terms of performance, a more uniform hot spot results in a more uniform beam shape. In addition, even with the additional two unnecessary posts, the PHI filaments will still slightly warp out of position over time. That in turn changes the sputter rate of the ion gun, which can result in inaccurate depth profile data unless you re-calibrate your sputter rate frequently.

In short, the RBD ion source filament assemblies outperform the PHI filaments, last longer, and the cost 75% less than a new ionizer. So the next time your 04-303 or 06-350 ion gun needs a new ion source filament assembly, send it to RBD for a rebuild!  You will save $$ and get the benefits of a longer lifetime ion source filament assembly and more stable sputter rates.

RBD carries exchange ion source filament assemblies so we can send you a rebuilt one and then you return your old one to us after you install the rebuilt ion source filament assembly.

Contact us for more information.

Bristol wrenches

bristol-set-screw

bristol-set-screw

Most of the set screws used on older PHI optics are the Bristol set screw type. The main advantage to this drive system is that almost all of the turning force is applied at right angles to the fastener spline face, which reduces the possibility of stripping the fastener. For this reason Bristol set screw are often used in softer, non-ferrous metals. Compared to an Allen drive, Bristol set screws are less likely to strip for the same amount of torque. Bristol set screws come with 4 flutes on the smaller sizes and 6 flutes on larger sizes.

When working on optics such as ion sources, analyzers, specimen stages and so on it is not uncommon to strip the set screws when removing them. A trick that can help prevent striping is to use a small amount of isopropanol as a lubricant before removing the set screws.

Some sources for replacement Bristol set screws are –

Fasteners & Metal Products Corporation

Cam-Tech Industries

Radax Industries

Note that these set screws are not cleaned for UHV and so you will need to run them through a cleaning process before using them for UHV optics applications.

bristol-wrench

bristol-wrench

When using Bristol set screws you need to use Bristol spline drivers. A source for Bristol spline drivers is –

Bristol Wrench Company, Inc  4 Hershey Drive, Ansonia, CT  06401

http://www.bristolwrench.com/

Note that Bristol Wrench no longer sells individual spline wrenches, only in sets of 10.  Good news is that you can get individual Bristol spline wrenches from McMaster Carr –

https://www.mcmaster.com/bristol-splines/