Amplifier Discriminator Tests

One of the troubleshooting techniques that can be used to test the counting circuitry of an auger or XPS system is called “tickling the brick”.   The term came about from when the early PHI systems used the Princeton Applied Research (PAR) 1120 amplifier discriminator for pulse count detection of electrons. The 1120 kind of looked like a brick, and so the term “tickling the brick” was born.

Basically, tickling the brick tests the brick (amplifier discriminator) and the counting circuitry.   This is one of the first troubleshooting steps that you would take if you are not getting any counts at all with your auger, XPS or SIMS system.

The 1120 amplifier discriminator is still in use on many PHI systems, as is its newer replacement, the 1182.  The 1120 and 1182 are used with single channel detectors.

In addition to the 1120 and 1182 amplifier discriminators, the “tickling the brick” concept can also be applied to PSD and MCD detectors on XPS systems.

The procedures for each amplifier discriminator type are described below.

Amplifier-Discriminator

1120/1182 Test Procedure

  1. Set up a alignment (can be any range) and select the 1120/1182 as the signal input. For most PHI systems using AugerScan software, that will be PC1 input in the hardware properties dialog box.
  2. Remove the bnc input that goes between the analyzer and the 1120/1182
  3. Turn off the analyzer control and electron multiplier supply as they are not needed.
  4. Start the alignment acquisition and then stick a piece of wire into the bnc input of the 1120/1182. This wire will act like an antenna and you should see between 20 million and 30 million CPS of noise.
  5. Remove the wire and the counts should go to zero.

If the counts behave as expected then both the 1120/1182 and the counting circuitry are working properly.

PSD Position Sensitive Detector Preamp Test Procedure

PSD

PSD

 

 

 

 

 

 

 

 

  1. Set up an alignment (can be any range) and select the PSD as the signal input. For most PHI systems using AugerScan software, that will be the PSD input in the hardware properties dialog box.
  2. Turn the card rack power OFF.
  3. Remove the bnc inputs to the PSD that goe between the analyzer and the PSD. These are the little white wires that go from Preamps A and B to the analyzer.
  4. Connect a test clip to the center pins of each of the little white wire bnc connectors.
  5. Start the alignment acquisition and then with your fingers touch the ends of both of the test clips that go to the center pins of each of the little white wire bnc connectors. The test clips will act like an antennae and you should see about 50K to 80K CPS of noise. Both clips need to be touched at the same time. NOTE: The clips go to the wires, not the BNC connectors on the analyzer.
  6. Remove the test clips and the counts should go to zero.

If the counts behave as expected then both the PSD preamp and the counting circuitry are working properly.

Make sure that the card rack power is OFF when you reconnect the PSD cables to the analyzer.

MCD Amplifier/Discriminator Test Procedure.

MCD

MCD

 

 

 

 

 

 

  1. Set up an alignment (can be any range) and select the MCD as the signal input. For most PHI systems using AugerScan software, that will be the MCD input in the XPS hardware properties dialog box.
  2. Turn the card rack power OFF.
  3. Unscrew the MCD amplifier discriminator from the analyzer flange and carefully remove it. NOTE: Use care when reinstalling it after the test procedure to make sure you do not bend any pins. Also make sure that the card rack power is OFF when you reinstall the amplifier discriminator on the analyzer flange
  4. Turn on the card rack power and start the alignment.
  5. There are 20 socket connectors visible on the end of the MCD amplifier discriminator. Each of those connectors (except for the middle 4) represents one of the 16 channels that are connected to the MCD detector inside the analyzer.
  6. Start the alignment acquisition and then stick a wire into each of the open pins on the end of the MCD preamp, one at a time. The test clips will act like an antennae and you should see about 20K to 80K CPS of noise when the clip is inserted.
  7. Remove the wire and the counts should go to zero.
  8. Repeat steps 6 and 7 for each of the other pins on the end of the MCD amplifier discriminator.

If the counts behave as expected then both the MCD amplifier/discriminator and the counting circuitry are working properly.

BCF Database for AES

Version 1.0 of the Backscattering-Correction-Factor Database for Auger Electron Spectroscopy program provided by NIST provides BSFs of homogeneous materials.

From the NIST website: This database provides values of backscattering correction factors (BCF) of homogeneous materials for quantitative surface analyses by Auger electron spectroscopy (AES). These BCFs are obtained from Monte Carlo simulations based on two models of electron transport in the material, a simplified model and an advanced model. One assumption for the former model is that the primary-electron beam is unchanged, in intensity, energy or direction, within the information depth for Auger-electron emission. This assumption becomes progressively less useful as the primary energy becomes closer to the core-level ionization energy for the relevant Auger transition or for increasing angles of incidence of the primary electrons.

BCFs can be calculated from both models so that users can readily ascertain the magnitudes of differences in BCFs from each model for materials and analysis conditions of interest. Analysts can readily specify the experimental conditions of interest (primary-beam energy, primary-beam angle of incidence, and, for the advanced model, analyzer-acceptance solid angle), the likely or estimated sample composition, the subshell of the element to be ionized, one of three available formulae for the inner-shell ionization cross section, and, for the advanced model, the Auger-electron transition of interest. The user can also select different numbers of trajectories in the Monte Carlo simulations so that tradeoffs can be made between calculation time and precision of the resulting BCF value. While simulations with the simplified model are generally faster than those with the advanced model, BCFs from the advanced model are considered more reliable. The results of a BCF calculation can be stored in a file for later use.

System Requirements: Personal computer operating on Windows 95, 98, NT, 2000, ME, XP, Vista, or 7, and hard disc space of at least 50 MB.System

Price: No Charge

You can download the program here – https://www-s.nist.gov/srd_online/index.cfm?fuseaction=home.main&productID=SRD154

 

Servicing a P-E TNBX Ion Pump Well

RBD Instruments provides all of the parts that you need to rebuild these TNBX ion pump wells.  Contact us for more information.  For best results, read this entire post and then look at the pictures at the bottom of the post.

The P-E (Perkin Elmer) TNBX style of Ion Pump well is found on most of the older PHI AES and ESCA systems. It is 12 inches in diameter and houses 8 ion pump elements. There is also a poppet valve that allows you to isolate the ion pump elements, keeping them under vacuum while venting the upper part of the chamber for maintenance. In this blog post we will show you how to access the ion pump elements for replacement or rebuild. The process includes instructions for disassembling and removing the poppet valve assembly, and removing all shielding and the TSP flange.

Important: Read through all procedures before beginning work to ensure you have the necessary equipment, personnel, and tools to complete the process.

Step 1: Remove the upper half of the chamber.

Equipment required: Chain hoist or engine hoist

  1. Turn off all controls and allow the ion gauge filament and the ion pump elements time to cool down.
  2. While the ion pump elements cool, remove all cables from the system.
  3. Prepare a space large enough to for the top half of the chamber. We recommend a piece of cardboard on the floor covered with UHV grade foil.
  4. Vent the chamber completely.
  5. Move the equipment you are going to use to remove the top of half of the system safely so that it’s within range of the chamber. We often use the chain hoist that we have at RBD Instruments. An engine hoist has proven to be a suitable alternative. Additional personnel may be helpful.
  6. Connect the chains from the equipment to the eyelets on the chamber and remove all bolts that hold the two halves together.
  7. Use the equipment to remove the upper half of the chamber. Additional personnel may be helpful to help guide the upper half of the chamber away from the bottom half.
  8. Place the upper half of the chamber in the space you created earlier. It will remain there while you are working on the pumps.

Step 2: Remove the Poppet valve to access the shielding.

Tools required: Two 7/16” wrenches.

  1. You should be able to look down into the chamber and see the poppet valve. You will likely see that the ion gauge that protrudes into the chamber above the poppet valve.
  2. Remove the ion gauge so that you do not damage it while you working inside of the pump well.
  3. Turn the poppet crank so that the poppet valve is all the way up.
  4. Grab hold of the poppet valve and rotate it counter-clockwise to unscrew it from the scissor jack.
  5. Rotate the poppet crank handle all of the way counter-clockwise. This crank handle is located on the side of the ion pump well.
  6. Remove the bolts that mount the poppet crank handle to the flange and rotate the crank handle and the crank housing together, counter-clockwise, until you can remove it from the scissor jack.
  7. Using two 7/16” wrenches, loosen and remove the mounting hardware for the scissor jack.
  8. With the mounting hardware out and the scissor jack free, work the scissor jack out of the ion pump well. This may take some maneuvering, but it will come out.

Step 3: Remove the TSP (Titanium Sublimation Pump) assembly and lower shielding.

Tools required: #1 flat head screw driver, #1 Phillip’s screw driver; 11/32, 3/8, 5/16 and ¼ wrenches.

  1. Remove the TSP cable from the TSP flange. If you’re lucky there is a screw-on connector for removing the TSP cable. The less convenient version has cap-head screws that fasten the wires to the flange feedthru’s. If you need to remove the wires, make sure that you mark the center “common” wire. This is the only wire that has a specific post that it must go to.
  2. With the wires removed you can now safely remove the TSP assembly.
  3. Remove the first piece of shielding. It is the shield that blocks titanium from the upper chamber and the poppet valve. The shielding is removed by loosening the mounting hardware. All shields and mounting hardware should be stored on a clean surface.
  4. Remove the circular cryo shield. This comes apart in sections for easier removal.
  5. You can now reach the high voltage conductor that runs from the high voltage feedthru to each of the ion pump elements.

Step 4: Ion Pump Element removal.

  1. Remove the high voltage conductor. It can either be one solid piece of ribbon or it can come in sections that can be removed one at a time. This conductor can be attached using a variety of different hardware, so be prepared with a variety of screw drivers and wrenches.
  2. Remove all of the ion pump elements once you’ve removed the high voltage conductor.
  3. We recommend that you remove the high voltage feedthru flange for inspection. It can sometimes get coated, and you also may find titanium flakes have worked their way up inside the flange. Once everything is removed and out of the way, you can clean the ion pump well.

Step 5: Cleaning and replacement of ion pump well and associated parts.

Equipment and tools required: Wire brush, shop vacuum

  1. Brush the walls, nooks, and crannies of the ion pump well to remove any build up titanium flakes.
  2. Vacuum out the ion pump well with a shop vacuum. Wrap the end of the hose in foil to prevent smearing the plastic hose with whatever might be on the ion pump well walls. It also prevents transferring whatever might be on the hose onto the walls of the ion pump well.
  3. Use isopropanol and Kim wipes to clean the inside of the ion pump well.
  4. Install the newly rebuilt or brand new ion pumps.
  5. Vacuum out the high voltage feedthru. If it’s obviously coated with titanium you will want to clean off the titanium using a sand blaster or some form of chemical etching process to prevent leakage current to ground.
  6. For the TSP assembly, RBD Instruments sells a variety of products to rebuild this assembly or you can even buy a complete assembly with or without exchange.
  7. Clean the shields using the above process described for cleaning the ion pump well (using the wire brush, vacuum, and isopropanol).
  8. Change the ion gauge filament, or if it’s in bad condition, replace the entire ion gauge. Keep in mind that RBD Instruments, Inc. sells ion gauge filaments and brand new assemblies as well.

Step 6: Reassembly

Tools required: 100 ft/lbs torque wrench

New parts required: New copper gaskets for all conflate flanges.

Note:    You can get the TNBX gasket here –

For Bolt Type Wire Seal Flanges, the wire diameter is 0.080”.
 
The smallest diameter we offer is Thermionics WSG Model #GW-140 (bolt type), ~12.962” I.D.
3118 Depot Road, Hayward, CA 94545
  1. Install the high voltage feedthru flange.
  2. Install the new or rebuilt ion pump elements.
  3. Install the Cryo shielding.
  4. Install the TSP assembly.
  5. Install the TSP shield.
  6. Install the scissor jack assembly and hardware.
  7. Install the poppet valve crank.
  8. Install the poppet valve. (First clean it if necessary, also remove and clean the O-ring, use a small amount of high vacuum grease on the O-ring before you put it back in the groove).
  9. Install the ion gauge.
  10. Install a 12.995” copper wire gasket on the flange that mates the lower half of the vacuum chamber to the upper half. When replacing the wire gasket it is important to exercise patience. The removal of the old one can be easy, but you will likely find that the new gasket is rather hard to put into place.  We have found that working the wire gasket around the flange with the handle end of a screw driver can work well to get the wire gasket to stretch into place onto the flange. You will see that it works a lot like putting a bike tire back on its rim, only in this case the material is copper and not rubber.
  11. With the lower half reassembled and the 12.995” copper wire gasket in place it’s time to position the top half of the chamber into place. When tightening down this flange it is good to go slowly and to use a torque wrench. The maximum torque on this flange should be 50 ft/lbs. We recommend starting at 25 ft/lbs and working your way to 45 ft/lbs in 10 ft/lbs increments. The last tightening will be 50 ft/lbs.
  12. That’s it! Time to start roughing the chamber down with your turbo and roughing pump.

    This slideshow requires JavaScript.