20-610 3kV Adjustment

Print Friendly, PDF & Email

The 20-610 high voltage gun supply used on PHI 600 and 660 scanning Auger systems provides the beam voltage, filament current and emission voltage to the Lab6 filament in the cylindrical mirror analyzer.

As part of the linearity adjustment process, the beam voltage is set to 3 keV so that the top of the elastic peak comes in at 3 keV.

The 3 keV elastic peak adjustment potentiometer in the 20-610 is R108.   The process is simple:

  1. Adjust the specimen stage Z axis for maximum counts and best shape of the peak during a 3 keV elastic peak alignment.
  2. Adjust the analyzer control gain so that the AES peaks come in at the correct location in a survey. Typically clean copper is used since it has both low and high energy peaks.
  3. Reacquire the elastic peak but do not move the specimen stage. If the elastic peak does not come it at 3 keV, move the elastic peak to 3 keV by adjusting R108 in the 20-610.

Sometimes there is not enough range of adjustment with the R108 potentiometer.  For those cases, this blog post will explain how to modify the PCB 100 board in the 20-610 in order to extend the 3 kV adjustment range.

The schematic below shows the R108 potentiometer circuit.

R108 Pot

The ends of R108 are connected to + and – 12V via two 49.9 K ohm resistors, R107 and R109. Adjusting R108 adds a small offset current to IC 103, which in turn changes the 3 kV output of the 20-610.

If there is not enough range with R108 then we need to change the balance between the + and – 12 volt supplies.   The way to do that is to remove R107 and replace it with a 100 k ohm trim potentiometer that is set to an initial value of 49.9 K ohms. 100 K ohm trim pots are available from any electronic component vendor (Digikey, Newark, Mouser….)

To prep the 100 K ohm trim potentiometer, the center wiper connection needs to be soldered to either end as shown below.

Bend center pin

With an ohmmeter, measure the resistance between the trim pot legs and adjust it for a resistance of 49.9 K ohms.

Turn off the 20-610 and on the back of the unit, unplug the main power cord, the remote program cable and the unscrew and remove the large HV cable connector. If there is a ground wire attached, remove it as well.

Place the 20-610 on a work bench and remove the 100 board. You will need to remove the board tie down bracket. Important! – Make sure that the 20-610 in unplugged when removing or installing the board tie down bracket as it is very close to some exposed wires that have voltage on them when the 20-610 is plugged in.

 

R100 board location in 20-610

remove bracket

 

 

 

Unsolder R107 and install the new trim pot on the back side of the 100 board.   You will need to bend the two pins on the trim pot so that the adjustment screw faces up.

R107 resistor

Reinstall the 100 board inside the 20-610.  Make sure that it is seated all the way into the mother board connector. Reinstall the board tie down bracket. Do not put the cover back on the 20-610.

Install the 20-610 into the electronics rack and reconnect the large HV cable (be sure to screw it in all the way), and the HV programming cable.  Make sure that the 20-610 main power switch is  OFF and then plug in the main power cord into the interlocked power strip.

Slide the 20-610 out enough so that you can get to R108 and the new trim pot.  Make sure that the 20-610 main power is OFF.

Remove the filament cap from the analyzer (3 screws) and connect a DVM and high voltage probe to either of the filament tabs.  The ground reference on the high voltage probe should be connected to the vacuum chamber.

 

***Caution, high voltage is present! Refer adjustment to qualified personnel ***

Turn on the 20-610 and using AugerMap and the normal turn on procedure, set the beam voltage to 3 kV.

***Caution, high voltage is present! Refer adjustment to qualified personnel ***

Center R108 and then adjust the new 100 K ohm trim pot so that the voltage on the filament cap is 3,000 volts.

Turn off the beam voltage in AugerMap and then turn off the 20-610.

Reconnect the filament cap to the analyzer.  Press down on the cap as you tighten the 3 screws that hold the cap onto the top of the CMA filament adjustment housing.

Turn the 20-610 back on and turn the beam voltage on and set it to 3 kV.  Bring up the filament current to the normal operating point (typically 1.3 amps)

Perform the AES calibration procedure and adjust R108 so that the elastic peak comes in at 3 keV.   When you are finished with the AES calibration, put the cover back on the 20-610 (front 2 cover screws only) and slide it back into the console.  The AES calibration procedure is listed below –

Auger energy calibration on 600 and 660 scanning Auger systems

This procedure requires sliding the 20-610 high voltage supply out and removing the cover to gain access to the beam voltage offset potentiometer, R108. Turn off the 20-610 when sliding it in out or in, and when removing or installing the cover.

Procedure:

  1. Load a sample of pure copper.
  2. If you are using AugerMap software, set the magnification to 10,000X and use the Area Scan mode to minimize sample topography effect on the Auger signal.
  3. Perform an elastic peak alignment and adjust the Z axis sample position to obtain maximum counts and best peak shape.
  4. Sputter the sample clean until no carbon or oxygen is present.
  5. Re-acquire the elastic peak to ensure that the sample is at the optimum position: highest counts and best peak shape. When the elastic peak is differentiated, the positive and negative excursions should be equal and symmetrical.
  6. From this point on, do not move the sample!
  7. With the beam voltage at 3kV, acquire a survey from 30eV to 1030eV, using .5eV/step, 50 ms/point.
  8. Differentiate the survey and check the peak positions against the correct values as listed in the PHI handbook or other reference. A typical value is 920eV for the high energy peak and 60eV for the low energy peak on copper.
  9. Note: If using AugerScan software, you can simply adjust the scale factor in the AES
  10. Acquire an alignment with a range of 900 to 940, .5eV/step, 15ms/point and do the adjustment in real time. For copper, set the n/e peak to approximately 917eV. When differentiated, the high energy Cu peak should be 920eV.
  11. Acquire another survey and check that the differentiated peak positions are correct. Document the results for future reference and file it in the system calibration log.
  12. Acquire another elastic peak, but do not move the sample!
  13. If the elastic peak is not centered at 3kV, then adjust R108 in the Bertan 20-610 High Voltage power supply to center the elastic peak.

Calibration is complete.

From this point on, every-time you set the elastic peak, the sample will be at the focal point of the analyzer (maximum signal and best shaped peak), and all of the Auger peaks will be in the correct positions.

Leave a Reply

Your email address will not be published. Required fields are marked *