{"id":523,"date":"2013-05-06T00:00:19","date_gmt":"2013-05-06T07:00:19","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=523"},"modified":"2014-11-06T09:02:02","modified_gmt":"2014-11-06T17:02:02","slug":"card-rack-power-supplies","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/","title":{"rendered":"Card rack power supplies"},"content":{"rendered":"<p>RBD Instruments recommends that you measure and also document the voltages on your card rack power supplies (located on the back right-hand electronic bay on most PHI systems) at least once per year. By monitoring these voltages you can notice trends that can indicate a problem with the card rack power supplies before those problems become evident by system performance issues. And, if you are having a high level problem with your system (such as no data), measuring the card rack power supplies is one of the first troubleshooting steps that you will need to perform.<\/p>\n<p>The two power supplies on the card rack door in the back of the electronics rack are:<\/p>\n<ol start=\"1\">\n<li>Pioneer Magnetics power supply (also referred to in PHI documentation as the OEM Power supply)<\/li>\n<li>Specialty power supply that provides +\/- 240 V DC, +\/- 18 V DC and -5.2 V DC voltages<\/li>\n<\/ol>\n<p>The Pioneer Magnetics supply outputs are shown below.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?ssl=1\"><img data-attachment-id=\"525\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/oem-power-supply-outputs\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=585%2C435&amp;ssl=1\" data-orig-size=\"585,435\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"oem power supply output test points\" data-image-description=\"&lt;p&gt;oem power supply output test points&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;oem power supply output test points&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=300%2C223&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=584%2C434&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-medium wp-image-525\" alt=\"oem power supply output test points\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=300%2C223&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs-300x223.jpg\" width=\"300\" height=\"223\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?resize=300%2C223&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?resize=403%2C300&amp;ssl=1 403w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?w=585&amp;ssl=1 585w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"525\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/oem-power-supply-outputs\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=585%2C435&amp;ssl=1\" data-orig-size=\"585,435\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"oem power supply output test points\" data-image-description=\"&lt;p&gt;oem power supply output test points&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;oem power supply output test points&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=300%2C223&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?fit=584%2C434&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-medium wp-image-525\" alt=\"oem power supply output test points\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?resize=300%2C223&#038;ssl=1\" width=\"300\" height=\"223\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?resize=300%2C223&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?resize=403%2C300&amp;ssl=1 403w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-outputs.jpg?w=585&amp;ssl=1 585w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<dl class=\"wp-caption alignleft\" id=\"attachment_525\" style=\"width: 310px;\">\n<dd class=\"wp-caption-dd\">oem power supply output test points<\/dd>\n<\/dl>\n<div id=\"attachment_524\" style=\"width: 310px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?ssl=1\"><img aria-describedby=\"caption-attachment-524\" data-attachment-id=\"524\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/pioneer-magnetics-power-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=567%2C425&amp;ssl=1\" data-orig-size=\"567,425\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"pioneer magnetics power supply\" data-image-description=\"&lt;p&gt;pioneer magnetics power supply&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;pioneer magnetics power supply&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=300%2C224&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=567%2C425&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-medium wp-image-524\" alt=\"pioneer magnetics power supply\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=300%2C224&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply-300x224.jpg\" width=\"300\" height=\"224\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?resize=300%2C224&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?resize=400%2C300&amp;ssl=1 400w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?w=567&amp;ssl=1 567w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-524\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"524\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/pioneer-magnetics-power-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=567%2C425&amp;ssl=1\" data-orig-size=\"567,425\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"pioneer magnetics power supply\" data-image-description=\"&lt;p&gt;pioneer magnetics power supply&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;pioneer magnetics power supply&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=300%2C224&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?fit=567%2C425&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-medium wp-image-524\" alt=\"pioneer magnetics power supply\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?resize=300%2C224&#038;ssl=1\" width=\"300\" height=\"224\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?resize=300%2C224&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?resize=400%2C300&amp;ssl=1 400w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/pioneer-magnetics-power-supply.jpg?w=567&amp;ssl=1 567w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> pioneer magnetics power supply<\/p><\/div>\n<div id=\"attachment_526\" style=\"width: 310px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?ssl=1\"><img aria-describedby=\"caption-attachment-526\" data-attachment-id=\"526\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/oem-power-supply-output-table\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=431%2C404&amp;ssl=1\" data-orig-size=\"431,404\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"oem power supply output table\" data-image-description=\"&lt;p&gt;oem power supply outputs&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;oem power supply outputs&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=300%2C281&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=431%2C404&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-medium wp-image-526\" alt=\"oem power supply outputs\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=300%2C281&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table-300x281.jpg\" width=\"300\" height=\"281\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?resize=300%2C281&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?resize=320%2C300&amp;ssl=1 320w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?w=431&amp;ssl=1 431w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-526\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"526\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/oem-power-supply-output-table\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=431%2C404&amp;ssl=1\" data-orig-size=\"431,404\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"oem power supply output table\" data-image-description=\"&lt;p&gt;oem power supply outputs&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;oem power supply outputs&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=300%2C281&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?fit=431%2C404&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-medium wp-image-526\" alt=\"oem power supply outputs\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?resize=300%2C281&#038;ssl=1\" width=\"300\" height=\"281\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?resize=300%2C281&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?resize=320%2C300&amp;ssl=1 320w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/oem-power-supply-output-table.jpg?w=431&amp;ssl=1 431w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> oem power supply outputs<\/p><\/div>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>The Specialty power supply outputs are shown below.<\/p>\n<div id=\"attachment_527\" style=\"width: 310px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?ssl=1\"><img aria-describedby=\"caption-attachment-527\" data-attachment-id=\"527\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/specialty-supply-outputs\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=593%2C318&amp;ssl=1\" data-orig-size=\"593,318\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Specialty supply outputs\" data-image-description=\"&lt;p&gt;Specialty supply outputs&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;Specialty supply outputs&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=300%2C160&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=584%2C313&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-medium wp-image-527\" alt=\"Specialty supply outputs\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=300%2C160&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs-300x160.jpg\" width=\"300\" height=\"160\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?resize=300%2C160&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?resize=500%2C268&amp;ssl=1 500w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?w=593&amp;ssl=1 593w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-527\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"527\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/specialty-supply-outputs\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=593%2C318&amp;ssl=1\" data-orig-size=\"593,318\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Specialty supply outputs\" data-image-description=\"&lt;p&gt;Specialty supply outputs&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;Specialty supply outputs&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=300%2C160&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?fit=584%2C313&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-medium wp-image-527\" alt=\"Specialty supply outputs\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?resize=300%2C160&#038;ssl=1\" width=\"300\" height=\"160\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?resize=300%2C160&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?resize=500%2C268&amp;ssl=1 500w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Specialty-supply-outputs.jpg?w=593&amp;ssl=1 593w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> Specialty supply outputs<\/p><\/div>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>RBD Instruments provides <a href=\"https:\/\/www.rbdinstruments.com\/services\/electronics-repair.html\">repair<\/a> and exchange services for both the Pioneer Magnetics OEM power supply and the PHI specialty supply.<\/p>\n<p>Have a qualified electronics technician measure and record the voltages at the points shown in the table below. Save this information with your system maintenance records as part of your Physical Electronics X-ray photoelectron or Auger electron spectrometer preventative maintenance program.<\/p>\n<div id=\"attachment_529\" style=\"width: 604px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?ssl=1\"><img aria-describedby=\"caption-attachment-529\" data-attachment-id=\"529\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/card-rack-power-supply-pm-form-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=705%2C283&amp;ssl=1\" data-orig-size=\"705,283\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Card rack power supply PM form\" data-image-description=\"&lt;p&gt;Card rack power supply Preventive Maintenance form&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;Card rack power supply Preventive Maintenance form&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=300%2C120&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=584%2C234&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-medium wp-image-529\" alt=\"Card rack power supply Preventive Maintenance form\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C233&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1-300x120.jpg\" width=\"584\" height=\"233\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?resize=300%2C120&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?resize=500%2C200&amp;ssl=1 500w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?w=705&amp;ssl=1 705w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-529\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"529\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/card-rack-power-supply-pm-form-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=705%2C283&amp;ssl=1\" data-orig-size=\"705,283\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Card rack power supply PM form\" data-image-description=\"&lt;p&gt;Card rack power supply Preventive Maintenance form&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;Card rack power supply Preventive Maintenance form&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=300%2C120&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?fit=584%2C234&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-medium wp-image-529\" alt=\"Card rack power supply Preventive Maintenance form\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?resize=584%2C233&#038;ssl=1\" width=\"584\" height=\"233\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?resize=300%2C120&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?resize=500%2C200&amp;ssl=1 500w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/Card-rack-power-supply-PM-form1.jpg?w=705&amp;ssl=1 705w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> Card rack power supply Preventive Maintenance form<\/p><\/div>\n","protected":false},"excerpt":{"rendered":"<p>RBD Instruments recommends that you measure and also document the voltages on your card rack power supplies (located on the back right-hand electronic bay on most PHI systems) at least once per year. By monitoring these voltages you can notice &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/card-rack-power-supplies\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":535,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[162],"tags":[74,75],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/05\/card-rack-power-supplies-mounted-on-door-in-back-of-electronics-console.jpg?fit=486%2C324&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-8r","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":3301,"url":"https:\/\/www.rbdinstruments.com\/blog\/testing-the-fuses-on-the-card-rack-power-supply-outputs\/","url_meta":{"origin":523,"position":0},"title":"Testing the fuses on the Card Rack power supply outputs","author":"Randy","date":"May 10, 2021","format":false,"excerpt":"The fuse strip located near the hinge on the OEM supply door routes the +5, +15, and -15 voltages from the OEM supply to the card rack unit motherboards. Each card rack unit has a dedicated section of fuse strip and specific fuse values. When troubleshooting problems with card rack\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1542,"url":"https:\/\/www.rbdinstruments.com\/blog\/11-155-power-supply-test\/","url_meta":{"origin":523,"position":1},"title":"11-155 power supply test","author":"Randy","date":"July 17, 2015","format":false,"excerpt":"This\u00a011-155 Voltage Testing PDF shows the test points to measure the voltages in the 11-155 power supply which is used on the older PHI 595 and 600 scanning auger systems. Procedure: \u00a0Slide the 11-155 power supply\u00a0 out so that the front of the unit is resting on the floor, and\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":1584,"url":"https:\/\/www.rbdinstruments.com\/blog\/dr11-data-entry\/","url_meta":{"origin":523,"position":2},"title":"DR11 data entry","author":"Randy","date":"October 11, 2015","format":false,"excerpt":"DR11 data entry for older PHI electronics Most of the digitally controlled power supplies on PHI card rack based and older stand alone power supply surface analysis systems used HEX commands to produce output voltages. By following the calibration procedures that are included with the manuals for the system the\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"147-diagnostic-dialog-box","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1420,"url":"https:\/\/www.rbdinstruments.com\/blog\/5600-xps-analyzer-voltages\/","url_meta":{"origin":523,"position":3},"title":"5600 XPS analyzer voltages","author":"Randy","date":"April 22, 2015","format":false,"excerpt":"One of the more common questions that we get when a PHI 5500 or 5600 XPS system loses the ability to collect data, is -\u00a0 \"how can we test the analyzer voltages?\"\u00a0\u00a0 This post will explain the procedure in detail. First though, the disclaimers - #1 This procedure involves measuring\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"5600 SCA filter box","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/04\/5600-SCA-filter-box.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/04\/5600-SCA-filter-box.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/04\/5600-SCA-filter-box.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":3199,"url":"https:\/\/www.rbdinstruments.com\/blog\/72-100-electron-multiplier-supply-notes\/","url_meta":{"origin":523,"position":4},"title":"72-100 Electron Multiplier Supply Notes","author":"Randy","date":"April 2, 2021","format":false,"excerpt":"Notes on to test and repair the 72-100 electron multiplier supply used on older PHI Auger systems.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":3675,"url":"https:\/\/www.rbdinstruments.com\/blog\/dgc-iii-filament-select-relay\/","url_meta":{"origin":523,"position":5},"title":"DGC III Filament Select Relay","author":"Randy","date":"February 9, 2023","format":false,"excerpt":"How to clean the filament select relay on a DGC III ion gauge controller.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"DGC III ion gauge controller","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/02\/Filament-relay.jpg?fit=900%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/02\/Filament-relay.jpg?fit=900%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/02\/Filament-relay.jpg?fit=900%2C1200&ssl=1&resize=700%2C400 2x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/523"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=523"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/523\/revisions"}],"predecessor-version":[{"id":1283,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/523\/revisions\/1283"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/535"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=523"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=523"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=523"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}