{"id":4154,"date":"2025-01-21T19:15:05","date_gmt":"2025-01-22T03:15:05","guid":{"rendered":"https:\/\/www.rbdinstruments.com\/blog\/?p=4154"},"modified":"2025-01-21T19:15:07","modified_gmt":"2025-01-22T03:15:07","slug":"contact-issue-with-mcd-ceramic-on-phi-5600-xps-system","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/","title":{"rendered":"Contact issue with MCD ceramic on PHI 5600 XPS system"},"content":{"rendered":"\n<p>This blog post will show you how to repair a damaged 16 channel MCD anode ceramic on a PHI 5600 X-ray Photoelectron Spectrometer system.<\/p>\n\n\n\n<p>The MCD anode ceramic is part of the MCD (multi channel detector) assembly.  The anode ceramic has 16 channels where electrons with slightly different energies are counted. The 16 channel detector offers a higher count rate as compared to the PSD (position sensitive detector) and single channel electron multipliers that were used on earlier XPS systems.<\/p>\n\n\n\n<p>There are surface mount resistors on the MCD anode ceramic that couple the positive (POS) voltage to the individual channel output ceramics.  One of these surface mount resistors in particular provides the pathway for the multiplier POS voltage to the channel (chevron) plates. The POS contact is on both sides of the ceramic and they are connected via a gold plated through hole.<\/p>\n\n\n\n<p>It is possible that over time (or when the chevron plates are replaced) that the gold plated pad can become electrically open in which case there would be no voltage applied to the chevron plates and, therefor, no XPS signal.<\/p>\n\n\n\n<p>The picture below shows the location of the POS voltage gold plated through hole.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?ssl=1\"><img data-attachment-id=\"4158\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250120_114125_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250120_114125_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-4158\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"4158\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250120_114125_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250120_114125_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-4158\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250120_114125_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/figure>\n\n\n\n<p>In this case, the gold was worn away from the gold plated through hole which resulted in an open circuit and no data. The picture below shows a close up of the gold plated through hole and the open area on the ceramic.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?ssl=1\"><img data-attachment-id=\"4159\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/missing-gold-on-contact\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1224%2C1024&amp;ssl=1\" data-orig-size=\"1224,1024\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"Missing-gold-on-contact\" data-image-description=\"\" data-image-caption=\"&lt;p&gt;open gold pad on MCD anode ceramic&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=300%2C251&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=584%2C489&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"489\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C489&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact-1024x857.jpg\" alt=\"\" class=\"lazy wp-image-4159\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=1024%2C857&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=300%2C251&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=768%2C643&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=359%2C300&amp;ssl=1 359w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?w=1224&amp;ssl=1 1224w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"4159\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/missing-gold-on-contact\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1224%2C1024&amp;ssl=1\" data-orig-size=\"1224,1024\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"Missing-gold-on-contact\" data-image-description=\"\" data-image-caption=\"&lt;p&gt;open gold pad on MCD anode ceramic&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=300%2C251&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=584%2C489&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"489\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=584%2C489&#038;ssl=1\" alt=\"\" class=\"wp-image-4159\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=1024%2C857&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=300%2C251&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=768%2C643&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?resize=359%2C300&amp;ssl=1 359w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?w=1224&amp;ssl=1 1224w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/figure>\n\n\n\n<p>Besides not getting any data this problem was also evident as the resistance between the POS and NEG connections was open instead of the normal 15 to 20 meg ohms.<\/p>\n\n\n\n<p>A new MCD anode ceramic costs about $700.00 and so I decided to first try and repair the contact using a thin strip of oxygen free copper foil <\/p>\n\n\n\n<p>First I used a ceramic cutting wheel to flatten out the out edge of the ceramic to make room for the copper foil. Then I cut the small piece of copper foil so that it would cover the pad by the resistor and then be able to bend over the other side of the ceramic so that it could contact the outer gold ring. I pressed the copper so that it would be as flat as possible.  Since there is a contact on both sides of the ceramic in the areas that the copper foil will cover, those contacts will hold the copper foil in place.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?ssl=1\"><img data-attachment-id=\"4160\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115558\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460558&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115558\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-4160\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"4160\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115558\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460558&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115558\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-4160\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115558-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?ssl=1\"><img data-attachment-id=\"4161\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115602\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460562&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115602\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-4161\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"4161\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115602\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460562&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115602\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-4161\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115602-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?ssl=1\"><img data-attachment-id=\"4162\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115630\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460590&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;64&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115630\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-4162\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"4162\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/20250121_115630\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=2560%2C2560&amp;ssl=1\" data-orig-size=\"2560,2560\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SM-G965U&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1737460590&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.3&quot;,&quot;iso&quot;:&quot;64&quot;,&quot;shutter_speed&quot;:&quot;0.016666666666667&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;1&quot;}\" data-image-title=\"20250121_115630\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-4162\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=1536%2C1536&amp;ssl=1 1536w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?resize=2048%2C2048&amp;ssl=1 2048w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?w=1168&amp;ssl=1 1168w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/20250121_115630-scaled.jpg?w=1752&amp;ssl=1 1752w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/figure>\n\n\n\n<p>After  reassembling the MCD assembly the multiplier POS to NEG resistance was 20 meg ohms, which is correct.<\/p>\n\n\n\n<p>This repair method works well if you do not have a replacement MCD ceramic handy.<\/p>\n\n\n\n<p>The PHI part number for the 16 channel anode ceramic is 619102 and the price as of January 2025 is $729.00 <\/p>\n\n\n\n<p>For more information on installing the Chevron plates into the MCD assembly refer to this blog post &#8211; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/5600-xps-system-mcd-channel-plate-replacement-procedure\/\">https:\/\/www.rbdinstruments.com\/blog\/5600-xps-system-mcd-channel-plate-replacement-procedure\/<\/a><\/p>\n\n\n\n<p><\/p>\n\n\n\n<p><\/p>\n\n\n\n<p><\/p>\n","protected":false},"excerpt":{"rendered":"<p>This blog post shows how to repair an open POS connector on a 16 channel MCD anode ceramic used in the MCD detector on PHI 5600 thru 5800 XPS systems. <a href=\"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":4159,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"This blog post shows how to repair the positive electrical contact on a 16 channel MCD anode ceramic.","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":false,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[164],"tags":[359],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1224%2C1024&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-150","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":3234,"url":"https:\/\/www.rbdinstruments.com\/blog\/5600-xps-system-mcd-channel-plate-replacement-procedure\/","url_meta":{"origin":4154,"position":0},"title":"5600 XPS system MCD channel plate replacement procedure.","author":"Randy","date":"April 28, 2021","format":false,"excerpt":"Procedure to install channel plates in a PHI 5600, 5700 or 5800 XPS system","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":3760,"url":"https:\/\/www.rbdinstruments.com\/blog\/electron-multipliers-for-older-phi-aes-and-xps-systems\/","url_meta":{"origin":4154,"position":1},"title":"Electron multipliers for older PHI AES and XPS systems","author":"Randy","date":"June 15, 2023","format":false,"excerpt":"Part numbers for electron multipliers for older Physical Electronics Auger and X-ray photoelectron analyzers.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":300,"url":"https:\/\/www.rbdinstruments.com\/blog\/phi-optics-repair-guidelines-2\/","url_meta":{"origin":4154,"position":2},"title":"PHI Optics Repair Guidelines","author":"Randy","date":"January 17, 2013","format":false,"excerpt":"This document contains information about optics repair methods, procedures and tricks that are useful when working on older PHI optics units such as cylindrical mirror analyzers, x-ray photoelectron spectrometers and sputter ion sources. General Optics Guidelines. Clean all tools with isopropanol or methanol and also degauss them if possible. Most\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/01\/10-155-electron-gun-detail.png?fit=619%2C568&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1211,"url":"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/","url_meta":{"origin":4154,"position":3},"title":"Multi channel detector continuity tests","author":"Randy","date":"September 11, 2014","format":false,"excerpt":"The multi channel detector (MCD) assembly found on Physical Electronics 5500 through 5800 X-ray photoelectron spectroscopy (XPS) systems comprises two pieces \u2013 the channel plate assembly and the connection flange. If you follow the procedure in the 10-360 spherical capacitive analyzer manual you can replace the channel plates (Chevron type)\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/09\/MCD-plates-oreintation1.jpg?resize=350%2C200","width":350,"height":200},"classes":[]},{"id":3449,"url":"https:\/\/www.rbdinstruments.com\/blog\/sca-charging-lens-elements\/","url_meta":{"origin":4154,"position":4},"title":"SCA charging lens elements","author":"Randy","date":"December 1, 2021","format":false,"excerpt":"This blog will describe the symptoms and solution for lens elements that charge up due to an oxidized graphite coating. Overview - The Physical Electronics SCA (Spherical Capaitive Analyzer) has a series of 2 or 4 lenses that focus electrons into the energy analyzer section of the SCA. The first\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1393,"url":"https:\/\/www.rbdinstruments.com\/blog\/amplifier-discriminator-tests\/","url_meta":{"origin":4154,"position":5},"title":"Amplifier Discriminator Tests","author":"Randy","date":"March 9, 2015","format":false,"excerpt":"One of the troubleshooting techniques that can be used to test the counting circuitry of an auger or XPS system is called \u201ctickling the brick\u201d.\u00a0\u00a0 The term came about from when the early PHI systems used the Princeton Applied Research (PAR) 1120 amplifier discriminator for pulse count detection of electrons.\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"amplifier-discriminator","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/03\/1120-amplifier-discriminator.jpg?fit=536%2C263&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/4154"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=4154"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/4154\/revisions"}],"predecessor-version":[{"id":4163,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/4154\/revisions\/4163"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/4159"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=4154"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=4154"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=4154"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}