{"id":3735,"date":"2023-05-04T07:46:54","date_gmt":"2023-05-04T14:46:54","guid":{"rendered":"https:\/\/www.rbdinstruments.com\/blog\/?p=3735"},"modified":"2023-05-04T07:48:51","modified_gmt":"2023-05-04T14:48:51","slug":"32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/","title":{"rendered":"32-095 and 32-096 X-ray source control faulty capacitor \u2013 urgent!"},"content":{"rendered":"\n<p>Recently, I have seen the same problem on several 32-095 and 32-096 X-ray source controls which are used on older Physical Electronics PHI X-ray photo electron spectroscopy systems.<\/p>\n\n\n\n<p>The issue is that C9, a 680 uF electrolytic capacitor blows out and the electrolytic material leaks out on the board.&nbsp; Left unattended, the electrolytic etches and oxidizes the traces on the board.<\/p>\n\n\n\n<p>If you have an older PHI XPS system that uses a 32-095 or 32-095 X-ray source control you should pull if out of the rack, remove the cover and inspect the board immediately.<\/p>\n\n\n\n<p>If corrosion is present, then remove the board and remove C9. &nbsp;Note the polarity of C9 as the + indicator on the board may be etched away. &nbsp;Then, carefully clean the corrosion from the board as best as you can.&nbsp;&nbsp; If in the shop I use some Alconox and let it sit on the board for a while, then rinse with DI water and let the board dry overnight.&nbsp;&nbsp; In the field I have used isopropanol or methanol and cotton swabs.&nbsp;&nbsp;&nbsp; Note that if the traces are corroded badly then they may come off the board as you clean it.&nbsp; If so, you will need to use some fine copper wire to rebuild the traces.<\/p>\n\n\n\n<p>Once the board is clean and dry, replace C9 with a new one.&nbsp; I will dig into this issue some more and try to determine why this problem occurs so often and come up with a permanent solution.&nbsp; In the meantime, I would recommend that the C9 capacitor be replaced every 5 years.<\/p>\n\n\n\n<p>The pictures below show where C9 is located on the control board and what the corrosion looks like.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?ssl=1\"><img data-attachment-id=\"3736\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122621_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122621_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3736\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3736\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122621_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122621_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3736\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><figcaption>C9 removed<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?ssl=1\"><img data-attachment-id=\"3737\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_111552_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_111552_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3737\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3737\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_111552_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_111552_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3737\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><figcaption>C9 location<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?ssl=1\"><img data-attachment-id=\"3738\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122456_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122456_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3738\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3738\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122456_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122456_resized\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3738\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><figcaption>C9 Neg towards connector<\/figcaption><\/figure>\n","protected":false},"excerpt":{"rendered":"<p>This blog post details a problem with 32-095 and 32-096 X-rat source controls related to a leaky capacitor. <a href=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":3736,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[164,162,160],"tags":[417,416,418,419],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-Yf","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":1911,"url":"https:\/\/www.rbdinstruments.com\/blog\/32-095-x-ray-source-control\/","url_meta":{"origin":3735,"position":0},"title":"32-095 X-ray source control","author":"Randy","date":"February 28, 2017","format":false,"excerpt":"When the 32-095 X-ray source control is not working properly there are some easy\u00a0 possible solutions that you can check before sending the unit to RBD Instruments Inc. for repair. If the problem is that one filament works but the other one does not, then you should check the fuses\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"32-095 local power supply board +5V","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/02\/32-095-local-power-supply-board-+5V.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/02\/32-095-local-power-supply-board-+5V.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/02\/32-095-local-power-supply-board-+5V.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":3746,"url":"https:\/\/www.rbdinstruments.com\/blog\/20-040-x-ray-source-power-supply-local-remote-switch\/","url_meta":{"origin":3735,"position":1},"title":"20-040 X-ray source power supply LOCAL REMOTE switch","author":"Randy","date":"May 30, 2023","format":false,"excerpt":"Procedure to bypass LOCAL \/ REMOTE switch on 20-040 X-ray source high voltage supply","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"20-040 front panel","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20-040-front-panel-scaled.jpg?fit=1200%2C599&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20-040-front-panel-scaled.jpg?fit=1200%2C599&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20-040-front-panel-scaled.jpg?fit=1200%2C599&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20-040-front-panel-scaled.jpg?fit=1200%2C599&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1591,"url":"https:\/\/www.rbdinstruments.com\/blog\/parts-that-fail-in-older-phi-electronic-units\/","url_meta":{"origin":3735,"position":2},"title":"Parts that fail in older PHI electronic units","author":"Randy","date":"October 31, 2015","format":false,"excerpt":"Common parts that fail in various older PHI electronic units This post lists common parts that fail in various PHI electronic units such as electron gun controls, analyzer controls and electron multiplier supplies. The part designations show an example of where a partial part may be used in a control.\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":4212,"url":"https:\/\/www.rbdinstruments.com\/blog\/16-020-050-heat-exchanger-shuts-off\/","url_meta":{"origin":3735,"position":3},"title":"16-020\/050 Heat Exchanger Shuts Off","author":"Randy","date":"February 5, 2026","format":false,"excerpt":"How to isolate the issue for when a 16-050 or 16-020 heat exchanger shuts down after a few seconds","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":4018,"url":"https:\/\/www.rbdinstruments.com\/blog\/16-050-heat-exchanger-flow-switch-replacement-procedure\/","url_meta":{"origin":3735,"position":4},"title":"16-050 Heat Exchanger Flow Switch replacement procedure.","author":"Randy","date":"April 2, 2024","format":false,"excerpt":"Procedure to install new flow switch into 16-050 heat exchanger used with Physical Electronics 5000 series XPS X-ray photo electron spectrometer","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"Remove 16-050 flow switch","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/04\/Install-new-flow-switch-1024x768-1.jpg?fit=1024%2C768&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/04\/Install-new-flow-switch-1024x768-1.jpg?fit=1024%2C768&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/04\/Install-new-flow-switch-1024x768-1.jpg?fit=1024%2C768&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1971,"url":"https:\/\/www.rbdinstruments.com\/blog\/x-ray-source-arcing\/","url_meta":{"origin":3735,"position":5},"title":"X-ray source arcing","author":"Randy","date":"May 8, 2017","format":false,"excerpt":"X-ray source arcing results in unstable XPS data and can also damage the x-ray source power supply or high voltage control. Typically X-ray source arcing is caused by contamination on the anode, a coated anode support ceramic (the football ceramic) or a loose filament. However one unseen cause of x-ray\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"burnt x-ray source connector","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=3735"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735\/revisions"}],"predecessor-version":[{"id":3741,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735\/revisions\/3741"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/3736"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=3735"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=3735"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=3735"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}