{"id":3735,"date":"2023-05-04T07:46:54","date_gmt":"2023-05-04T14:46:54","guid":{"rendered":"https:\/\/www.rbdinstruments.com\/blog\/?p=3735"},"modified":"2023-05-04T07:48:51","modified_gmt":"2023-05-04T14:48:51","slug":"32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/","title":{"rendered":"32-095 and 32-096 X-ray source control faulty capacitor \u2013 urgent!"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Recently, I have seen the same problem on several 32-095 and 32-096 X-ray source controls which are used on older Physical Electronics PHI X-ray photo electron spectroscopy systems.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The issue is that C9, a 680 uF electrolytic capacitor blows out and the electrolytic material leaks out on the board.&nbsp; Left unattended, the electrolytic etches and oxidizes the traces on the board.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you have an older PHI XPS system that uses a 32-095 or 32-095 X-ray source control you should pull if out of the rack, remove the cover and inspect the board immediately.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If corrosion is present, then remove the board and remove C9. &nbsp;Note the polarity of C9 as the + indicator on the board may be etched away. &nbsp;Then, carefully clean the corrosion from the board as best as you can.&nbsp;&nbsp; If in the shop I use some Alconox and let it sit on the board for a while, then rinse with DI water and let the board dry overnight.&nbsp;&nbsp; In the field I have used isopropanol or methanol and cotton swabs.&nbsp;&nbsp;&nbsp; Note that if the traces are corroded badly then they may come off the board as you clean it.&nbsp; If so, you will need to use some fine copper wire to rebuild the traces.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Once the board is clean and dry, replace C9 with a new one.&nbsp; I will dig into this issue some more and try to determine why this problem occurs so often and come up with a permanent solution.&nbsp; In the meantime, I would recommend that the C9 capacitor be replaced every 5 years.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The pictures below show where C9 is located on the control board and what the corrosion looks like.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?ssl=1\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3736\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122621_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122621_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3736\"\/><noscript><img data-recalc-dims=\"1\" loading=\"lazy\" decoding=\"async\" width=\"584\" height=\"584\" data-attachment-id=\"3736\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122621_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122621_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3736\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><\/a><figcaption>C9 removed<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?ssl=1\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3737\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_111552_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_111552_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3737\"\/><noscript><img data-recalc-dims=\"1\" loading=\"lazy\" decoding=\"async\" width=\"584\" height=\"584\" data-attachment-id=\"3737\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_111552_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_111552_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3737\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_111552_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><\/a><figcaption>C9 location<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?ssl=1\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3738\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122456_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122456_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized-1024x1024.jpg\" alt=\"\" class=\"lazy wp-image-3738\"\/><noscript><img data-recalc-dims=\"1\" loading=\"lazy\" decoding=\"async\" width=\"584\" height=\"584\" data-attachment-id=\"3738\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/20230503_122456_resized\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"20230503_122456_resized\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=584%2C584&#038;ssl=1\" alt=\"\" class=\"wp-image-3738\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122456_resized.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><\/a><figcaption>C9 Neg towards connector<\/figcaption><\/figure>\n","protected":false},"excerpt":{"rendered":"<p>This blog post details a problem with 32-095 and 32-096 X-rat source controls related to a leaky capacitor. <a href=\"https:\/\/www.rbdinstruments.com\/blog\/32-095-and-32-096-x-ray-source-control-faulty-capacitor-urgent\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":3736,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_feature_clip_id":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2},"jetpack_post_was_ever_published":false},"categories":[164,162,160],"tags":[417,416,418,419],"class_list":["post-3735","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-general-optics-and-vacuum","category-operation-and-calibration-procedures","category-x-ray-sources","tag-32-096","tag-416","tag-blown-capacitor","tag-c-9"],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/05\/20230503_122621_resized.jpg?fit=1512%2C1512&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-Yf","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=3735"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735\/revisions"}],"predecessor-version":[{"id":3741,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3735\/revisions\/3741"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/3736"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=3735"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=3735"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=3735"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}