{"id":3221,"date":"2021-04-13T08:00:40","date_gmt":"2021-04-13T15:00:40","guid":{"rendered":"https:\/\/www.rbdinstruments.com\/blog\/?p=3221"},"modified":"2021-04-12T20:02:49","modified_gmt":"2021-04-13T03:02:49","slug":"72-030-neutralizer-current-limit-procedure","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/","title":{"rendered":"72-030 Neutralizer Current limit procedure"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">After you <a href=\"https:\/\/www.rbdinstruments.com\/blog\/neutralizer-filament-replacement-procedure\/\">replace the filament<\/a> in the 04-085 or 04-090 neutralizer you may need to readjust the current limit on the 72-030 neutralizer board.  The 72-030 provides the filament current and bias voltage to the neutralizer. This procedure shows you how to set the current limit on the 72-030 so that you generate enough electrons to neutralize the sample and also maximize the filament lifetime.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Tools Needed:<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">HV card rack extender board<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Voltmeter with 2 clip leads<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Potentiometer adjustment tool<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>72-030 Calibration procedure<\/strong><\/p>\n\n\n\n<ol class=\"wp-block-list\"><li>Turn off the card rack power<\/li><li>Extend the 72-030 neutralizer card (located in the card rack)<\/li><li>Connect the DVM across R59 \/ E6<\/li><li>Make sure that the filament cable is connected to the neutralizer on the system<\/li><li>Turn on the card rack power<\/li><li>Turn on the 147 PC Interface unit<\/li><li>Open AugerScan software<\/li><li>Turn on the neutralizer in AugerScan <\/li><li>Monitor the target current with a picoammeter and a +90V bias<\/li><li>Turn up the Emission in the Augerscan Neutralizer dialog box to 20 to 23 mA while monitoring the DC voltage in millivolts across R59 \/ E6.<\/li><li>There should be approximately 34 mV across R59 \/ E6 and you should also be able to see the end of the neutralizer glowing orange to white  in the vacuum chamber.<\/li><li>Set the Emission to 25 mA and if necessary adjust R48 \/ B6 so that you have 2 to 3 uA of target current.\u00a0\u00a0 Do not adjust R48 \/ B6 to exceed 36mV across R59 \/ E6.   36mV corresponds to 3.6 amps of filament current.  The lower the filament current, the longer the filament lifetime. <\/li><li>Turn off the card rack power and reinstall the 72-030 neutralizer card in the card rack. <\/li><\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Ideally, adjusting R48 for a value that provides a few uA of target current and with less than 36mV across R59 \/ E6 will result in the longest filament lifetime.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3224\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/72-030-board-extended\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"72-030-board-extended\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3224\"\/><noscript><img loading=\"lazy\" decoding=\"async\" width=\"1512\" height=\"1512\" data-attachment-id=\"3224\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/72-030-board-extended\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"72-030-board-extended\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3224\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>72-030 board extended<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3225\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/r59-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"R59-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3225\"\/><noscript><img loading=\"lazy\" decoding=\"async\" width=\"1512\" height=\"1512\" data-attachment-id=\"3225\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/r59-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"R59-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3225\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>R59 current sense resistor<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3226\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/measure-across-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=1280%2C1280&amp;ssl=1\" data-orig-size=\"1280,1280\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"measure-across-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3226\"\/><noscript><img loading=\"lazy\" decoding=\"async\" width=\"1280\" height=\"1280\" data-attachment-id=\"3226\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/measure-across-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=1280%2C1280&amp;ssl=1\" data-orig-size=\"1280,1280\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"measure-across-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3226\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1280&amp;ssl=1 1280w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>Measure voltage across R59     34mV corresponds to 3.4 amps of filament current<\/figcaption><\/figure>\n\n\n<p><br>\n<br>\n<!--StartFragment--><\/p>\n\n\n<p class=\"wp-block-paragraph\"><strong>72-030 Neutralizer bias voltage notes<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">There are two revisions of 72-030 neutralizer boards.\u00a0 The older original 72-030 has a bias voltage range of 0 to -12V.\u00a0\u00a0 The newer 72-030s have a bias voltage range of 0 to -24V.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">You can tell the difference by two things.\u00a0 First of all, U6 is a OP07 op amp on the older 72-030s, and it is a LM344H high voltage op amp on the newer 72-030s.\u00a0 Secondly, there is a 5V to 15V power convertor on the newer boards, the older 72-030s do not have that power converter.<\/p>\n\n\n<p><!--EndFragment--><br \/><br \/><\/p>\n\n\n<figure class=\"wp-block-image\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3227\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/12v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"12V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3227\"\/><noscript><img loading=\"lazy\" decoding=\"async\" width=\"1512\" height=\"1512\" data-attachment-id=\"3227\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/12v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"12V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3227\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>12V bias 72-030<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-recalc-dims=\"1\" decoding=\"async\" data-attachment-id=\"3228\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/24-v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"24-V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?w=584&#038;ssl=1\" data-original=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3228\"\/><noscript><img loading=\"lazy\" decoding=\"async\" width=\"1512\" height=\"1512\" data-attachment-id=\"3228\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/24-v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"24-V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3228\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"auto, (max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>24V bias 72-030<\/figcaption><\/figure>\n","protected":false},"excerpt":{"rendered":"<p>This procedure shows how to adjust the filament current on the 72-030 neutralizer controller in order to generate sufficient electrons for charge neuralization.  <a href=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":3224,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_feature_clip_id":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2},"jetpack_post_was_ever_published":false},"categories":[162],"tags":[357,358],"class_list":["post-3221","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-operation-and-calibration-procedures","tag-72-030","tag-neutralizer-controller"],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-PX","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=3221"}],"version-history":[{"count":3,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221\/revisions"}],"predecessor-version":[{"id":3298,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221\/revisions\/3298"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/3224"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=3221"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=3221"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=3221"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}