{"id":3221,"date":"2021-04-13T08:00:40","date_gmt":"2021-04-13T15:00:40","guid":{"rendered":"https:\/\/www.rbdinstruments.com\/blog\/?p=3221"},"modified":"2021-04-12T20:02:49","modified_gmt":"2021-04-13T03:02:49","slug":"72-030-neutralizer-current-limit-procedure","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/","title":{"rendered":"72-030 Neutralizer Current limit procedure"},"content":{"rendered":"\n<p>After you <a href=\"https:\/\/www.rbdinstruments.com\/blog\/neutralizer-filament-replacement-procedure\/\">replace the filament<\/a> in the 04-085 or 04-090 neutralizer you may need to readjust the current limit on the 72-030 neutralizer board.  The 72-030 provides the filament current and bias voltage to the neutralizer. This procedure shows you how to set the current limit on the 72-030 so that you generate enough electrons to neutralize the sample and also maximize the filament lifetime.<\/p>\n\n\n\n<p><strong>Tools Needed:<\/strong><\/p>\n\n\n\n<p>HV card rack extender board<\/p>\n\n\n\n<p>Voltmeter with 2 clip leads<\/p>\n\n\n\n<p>Potentiometer adjustment tool<\/p>\n\n\n\n<p><strong>72-030 Calibration procedure<\/strong><\/p>\n\n\n\n<ol><li>Turn off the card rack power<\/li><li>Extend the 72-030 neutralizer card (located in the card rack)<\/li><li>Connect the DVM across R59 \/ E6<\/li><li>Make sure that the filament cable is connected to the neutralizer on the system<\/li><li>Turn on the card rack power<\/li><li>Turn on the 147 PC Interface unit<\/li><li>Open AugerScan software<\/li><li>Turn on the neutralizer in AugerScan <\/li><li>Monitor the target current with a picoammeter and a +90V bias<\/li><li>Turn up the Emission in the Augerscan Neutralizer dialog box to 20 to 23 mA while monitoring the DC voltage in millivolts across R59 \/ E6.<\/li><li>There should be approximately 34 mV across R59 \/ E6 and you should also be able to see the end of the neutralizer glowing orange to white  in the vacuum chamber.<\/li><li>Set the Emission to 25 mA and if necessary adjust R48 \/ B6 so that you have 2 to 3 uA of target current.\u00a0\u00a0 Do not adjust R48 \/ B6 to exceed 36mV across R59 \/ E6.   36mV corresponds to 3.6 amps of filament current.  The lower the filament current, the longer the filament lifetime. <\/li><li>Turn off the card rack power and reinstall the 72-030 neutralizer card in the card rack. <\/li><\/ol>\n\n\n\n<p>Ideally, adjusting R48 for a value that provides a few uA of target current and with less than 36mV across R59 \/ E6 will result in the longest filament lifetime.<\/p>\n\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3224\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/72-030-board-extended\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"72-030-board-extended\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3224\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3224\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/72-030-board-extended\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"72-030-board-extended\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3224\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>72-030 board extended<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3225\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/r59-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"R59-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3225\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3225\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/r59-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"R59-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3225\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/R59-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>R59 current sense resistor<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3226\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/measure-across-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=1280%2C1280&amp;ssl=1\" data-orig-size=\"1280,1280\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"measure-across-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3226\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1280&amp;ssl=1 1280w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3226\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/measure-across-current-sense-resistor\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=1280%2C1280&amp;ssl=1\" data-orig-size=\"1280,1280\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"measure-across-current-sense-resistor\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1280\" height=\"1280\" src=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3226\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1280&amp;ssl=1 1280w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/measure-across-current-sense-resistor.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>Measure voltage across R59     34mV corresponds to 3.4 amps of filament current<\/figcaption><\/figure>\n\n\n<p><br>\n<br>\n<!--StartFragment--><\/p>\n\n\n<p><strong>72-030 Neutralizer bias voltage notes<\/strong><\/p>\n\n\n\n<p>There are two revisions of 72-030 neutralizer boards.\u00a0 The older original 72-030 has a bias voltage range of 0 to -12V.\u00a0\u00a0 The newer 72-030s have a bias voltage range of 0 to -24V.<\/p>\n\n\n\n<p>You can tell the difference by two things.\u00a0 First of all, U6 is a OP07 op amp on the older 72-030s, and it is a LM344H high voltage op amp on the newer 72-030s.\u00a0 Secondly, there is a 5V to 15V power convertor on the newer boards, the older 72-030s do not have that power converter.<\/p>\n\n\n<p><!--EndFragment--><br \/><br \/><\/p>\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3227\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/12v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"12V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3227\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3227\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/12v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"12V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i2.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3227\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/12V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>12V bias 72-030<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3228\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/24-v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"24-V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3228\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3228\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/24-v-bias-supply\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"24-V-bias-supply\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3228\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/24-V-bias-supply.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>24V bias 72-030<\/figcaption><\/figure>\n","protected":false},"excerpt":{"rendered":"<p>This procedure shows how to adjust the filament current on the 72-030 neutralizer controller in order to generate sufficient electrons for charge neuralization.  <a href=\"https:\/\/www.rbdinstruments.com\/blog\/72-030-neutralizer-current-limit-procedure\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":3224,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[162],"tags":[357,358],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/72-030-board-extended.jpg?fit=1512%2C1512&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-PX","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":755,"url":"https:\/\/www.rbdinstruments.com\/blog\/neutralizer-filament-replacement-procedure\/","url_meta":{"origin":3221,"position":0},"title":"Neutralizer filament replacement procedure","author":"Randy","date":"September 2, 2013","format":false,"excerpt":"This post will detail the neutralizer filament replacement procedure for the Physical Electronics (PHI) 04-085 and 04-090 neutralizers that are commonly found on many PHI X-ray photoelectron spectrometers and Static SIMS systems. Before you can replace the filament you need a new filament. We also recommend having the deflection ceramic\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/09\/04085-neutralizer.jpg?fit=486%2C324&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1553,"url":"https:\/\/www.rbdinstruments.com\/blog\/xps-pm-procedure\/","url_meta":{"origin":3221,"position":1},"title":"XPS PM Procedure","author":"Randy","date":"August 12, 2015","format":false,"excerpt":"5400\/5500\/5600 System XPS PM procedure Preventive Maintenance Guide This post lists the common tasks that are completed as part of a preventative maintenance on a PHI 5400\/5500\/5600 XPS X-ray photo electron spectrometer. XPS PM Procedure - OPTICS *Discuss System performance and issues with system operator prior to servicing. *Inspect system,\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"5600 XPS system with mono","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/04\/5600-XPS-system-with-monochromator.jpg?fit=500%2C333&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":4064,"url":"https:\/\/www.rbdinstruments.com\/blog\/a-numbers-for-phi-units\/","url_meta":{"origin":3221,"position":2},"title":"&#8220;A&#8221; Numbers for PHI units","author":"Randy","date":"September 9, 2024","format":false,"excerpt":"A numbers for older PHI surface analysis systems control units.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1597,"url":"https:\/\/www.rbdinstruments.com\/blog\/pm-sequence-on-xps-and-aes-systems\/","url_meta":{"origin":3221,"position":3},"title":"PM sequence on XPS and AES systems","author":"Randy","date":"December 6, 2015","format":false,"excerpt":"Performing regular Preventive Maintenance on surface analysis systems such as X-ray Photo Electron spectrometers (XPS) and Auger Electron spectrometers (AES) is a very important step in keeping the systems functioning properly and reliably. But performing the necessary steps in the correct PM sequence will make sure that you maximize the\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"electronics console air filters","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2012\/11\/GMBH-filters.jpg?fit=704%2C470&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2012\/11\/GMBH-filters.jpg?fit=704%2C470&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2012\/11\/GMBH-filters.jpg?fit=704%2C470&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1629,"url":"https:\/\/www.rbdinstruments.com\/blog\/5400-ls-operating-procedure\/","url_meta":{"origin":3221,"position":4},"title":"5400 LS operating procedure","author":"Randy","date":"February 9, 2016","format":false,"excerpt":"The following 5400 LS operating procedures provide details of many aspects of how to operate a 5000 LS XPS system. PHI did not make very many of the 5400 LS (large sample) systems but the ones that are still in operation are real work horses. On a side note, I\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":1584,"url":"https:\/\/www.rbdinstruments.com\/blog\/dr11-data-entry\/","url_meta":{"origin":3221,"position":5},"title":"DR11 data entry","author":"Randy","date":"October 11, 2015","format":false,"excerpt":"DR11 data entry for older PHI electronics Most of the digitally controlled power supplies on PHI card rack based and older stand alone power supply surface analysis systems used HEX commands to produce output voltages. By following the calibration procedures that are included with the manuals for the system the\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"147-diagnostic-dialog-box","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/10\/147-diagnostic-dialog-box.jpg?fit=713%2C537&ssl=1&resize=700%2C400 2x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=3221"}],"version-history":[{"count":3,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221\/revisions"}],"predecessor-version":[{"id":3298,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/3221\/revisions\/3298"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/3224"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=3221"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=3221"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=3221"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}