{"id":1837,"date":"2016-12-07T16:04:06","date_gmt":"2016-12-08T00:04:06","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=1837"},"modified":"2016-12-07T16:15:27","modified_gmt":"2016-12-08T00:15:27","slug":"calibrating-the-simsii-rf-generator","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/","title":{"rendered":"Calibrating the SIMSII RF Generator"},"content":{"rendered":"<p><strong>Calibrating the SIMSII RF Generator<\/strong><\/p>\n<p>This post shows you how to calibrate the Balzers QMH RF generators that are used on the PHI SIMSII and 6300\/6600 series SIMS systems.<\/p>\n<p><strong>1.<\/strong> Load in a Molybdenum sample. Note that the sample mount covers on PHI sample mounts are made out of Molybdenum so if you have one of those on your sample you can use the cover.<\/p>\n<p><strong>2.<\/strong> Monitor the voltage on the two recessed banana jacks on the front of the RF generator with a DVM set to DC millivolts.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?ssl=1\"><img data-attachment-id=\"1840\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/rf-generator-tuning-test-points-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=648%2C972&amp;ssl=1\" data-orig-size=\"648,972\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;Canon EOS DIGITAL REBEL XTi&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1480666834&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;21&quot;,&quot;iso&quot;:&quot;400&quot;,&quot;shutter_speed&quot;:&quot;0.0166666666667&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"RF generator tuning test points (2)\" data-image-description=\"&lt;p&gt;tuning_test_points&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=200%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=584%2C876&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1840\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=400%2C600&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg\" alt=\"tuning_test_points\" width=\"400\" height=\"600\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?w=648&amp;ssl=1 648w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?resize=200%2C300&amp;ssl=1 200w\" sizes=\"(max-width: 400px) 100vw, 400px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1840\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/rf-generator-tuning-test-points-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=648%2C972&amp;ssl=1\" data-orig-size=\"648,972\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;4&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;Canon EOS DIGITAL REBEL XTi&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1480666834&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;21&quot;,&quot;iso&quot;:&quot;400&quot;,&quot;shutter_speed&quot;:&quot;0.0166666666667&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"RF generator tuning test points (2)\" data-image-description=\"&lt;p&gt;tuning_test_points&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=200%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?fit=584%2C876&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1840\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?resize=400%2C600&#038;ssl=1\" alt=\"tuning_test_points\" width=\"400\" height=\"600\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?w=648&amp;ssl=1 648w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/RF-generator-tuning-test-points-2.jpg?resize=200%2C300&amp;ssl=1 200w\" sizes=\"(max-width: 400px) 100vw, 400px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p><strong>3.<\/strong>\u00a0 In AugerScan set up a SIMS alignment with a Peak AMU of 10 and a range of zero. \u00a0Note that for this adjustment the ion beam does not need to be on.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?ssl=1\"><img data-attachment-id=\"1838\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/10_amu_alignment\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=300%2C387&amp;ssl=1\" data-orig-size=\"300,387\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"10_AMU_Alignment\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=232%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=300%2C387&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1838 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=300%2C387&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png\" alt=\"sims_alignment_dialog_box\" width=\"300\" height=\"387\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?w=300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?resize=232%2C300&amp;ssl=1 232w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1838\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/10_amu_alignment\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=300%2C387&amp;ssl=1\" data-orig-size=\"300,387\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"10_AMU_Alignment\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=232%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?fit=300%2C387&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1838 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?resize=300%2C387&#038;ssl=1\" alt=\"sims_alignment_dialog_box\" width=\"300\" height=\"387\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?w=300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10_AMU_Alignment.png?resize=232%2C300&amp;ssl=1 232w\" sizes=\"(max-width: 300px) 100vw, 300px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p><strong>4.<\/strong> Acquire the alignment and adjust the Tuning capacitor on the back side of the RF generator for the minimum voltage.\u00a0\u00a0 The capacitor is a tuning type like what was used in old radios. The shaft of the capacitor is attached to a wheel with holes in it. To adjust the capacitor you put a small thin screwdriver or an Allen wrench into one of the holes and move the capacitor wheel left and right. There is a little bit of hysteresis in the adjustment so you will need to go back and forth a little bit to make sure that you are at the minimum voltage value.<\/p>\n<p>The DC voltage will minimize at about 10 to 20mV depending on the RF generator AM range and the size of the quadrupole.<\/p>\n<p>Once you have minimized the DC voltage stop the alignment. Change the AMU to 50 and repeat step 4.<\/p>\n<p><strong>5.<\/strong> Next change the AMU in the alignment to 200 AMU and adjust for minimum voltage. 200 AMU is as high as you will need to go as most of the RF generators only go up to a maximum of 250 AMU. But even if you have one that goes up to 500 or even 1000AMU, the effective signal above 200 AMU is minimal and so I think that it makes more sense to optimize the tuning over the effective range of the RF generator rather than the absolute maximum AMU. \u00a0\u00a0 The voltage at 200 amu will be about 400 to 600mV depending RF generator AMU range and the quad diameter.\u00a0 The important thing is to minimize the voltage at 200 AMU.<\/p>\n<p>After tuning the next step will be to adjust the peak positions.<\/p>\n<p><strong>6.<\/strong> The ion beam needs to be set up for normal operation and the raster turned OFF. Open the O2 leak valve on the SIMS II energy analyzer (top of SIMSII) and set it so that the vacuum in the chamber is 2 X 10-7 Torr. There is some hysteresis in the gas line so be very careful when adjusting this valve. My technique is to slowly open the valve and as soon as I see the pressure in the chamber starting to come up I close the valve \u00bc turn and wait for the pressure to stabilize before further adjustment.<\/p>\n<p><strong>7.<\/strong> Set up an alignment in POS SIMS with an AMU of 16 and a range of 3. Set the time per step to 50ms.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?ssl=1\"><img data-attachment-id=\"1843\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/sims-ii-adjustment-pots\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=717%2C755&amp;ssl=1\" data-orig-size=\"717,755\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"SIMS II adjustment pots\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=284%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=584%2C615&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1843 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C615&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg\" alt=\"sims_adjustment_pots\" width=\"584\" height=\"615\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?w=717&amp;ssl=1 717w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?resize=284%2C300&amp;ssl=1 284w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1843\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/sims-ii-adjustment-pots\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=717%2C755&amp;ssl=1\" data-orig-size=\"717,755\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"SIMS II adjustment pots\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=284%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=584%2C615&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1843 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?resize=584%2C615&#038;ssl=1\" alt=\"sims_adjustment_pots\" width=\"584\" height=\"615\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?w=717&amp;ssl=1 717w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?resize=284%2C300&amp;ssl=1 284w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p><strong>\u00a08.<\/strong> Acquire the alignment and adjust the Cal LOW potentiometer on the RF generator so that the 16 AMU O2 peak is centered. Note that for this adjustment the ion beam needs to be on. CCW moves peak up.<\/p>\n<p><strong>9.<\/strong> Change the AMU to 98 and a range of 3.\u00a0\u00a0 Adjust the Cal HIGH potentiometer on the RF generator so that the 98 AMU moly peak is centered. The 98 Moly peak will be the highest peak in the group of peaks between 92 and 100 AMU. \u00a0CCW moves peak up.<\/p>\n<p><strong>10.<\/strong> Recheck the 16 AMU peak and readjust if necessary. If you adjust the 16 AMU peak then you will also need to adjust the 98 AMU peak as well.<\/p>\n<p>Next we will adjust the resolution linearity. The resolution is adjusted by a combination of the Resolution number in the software and also the actual resolution fine potentiometer on the RF generator. We will adjust the resolution for what is called the 10% value. The 10% value represents the ratio between the 95 and 96 Moly peaks and when viewed in the log scale is one order of magnitude on the graph. We further will want to adjust the resolution linearity so that the peaks from 90 to 100 AMU and also 185 to 195 AMU on Moly are both showing approximately a 10% peak to valley ratio, as shown in the survey below.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?ssl=1\"><img data-attachment-id=\"1844\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/sims-linearity-on-moly-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=686%2C458&amp;ssl=1\" data-orig-size=\"686,458\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Sims linearity on Moly (2)\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=300%2C200&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=584%2C390&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1844 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C390&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg\" alt=\"10_percent_linearity\" width=\"584\" height=\"390\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?w=686&amp;ssl=1 686w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?resize=300%2C200&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?resize=449%2C300&amp;ssl=1 449w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1844\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/sims-linearity-on-moly-2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=686%2C458&amp;ssl=1\" data-orig-size=\"686,458\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Sims linearity on Moly (2)\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=300%2C200&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?fit=584%2C390&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1844 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?resize=584%2C390&#038;ssl=1\" alt=\"10_percent_linearity\" width=\"584\" height=\"390\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?w=686&amp;ssl=1 686w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?resize=300%2C200&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/Sims-linearity-on-Moly-2.jpg?resize=449%2C300&amp;ssl=1 449w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>The resolution number in the software acts like a resolution offset \u2013 it affects the entire range of AMU. The resolution gain potentiometer on the RF generator acts like a gain, it has more effect on the high AMU than the low AMU. By selecting the correct Resolution number in the software and adjusting the resolution Fine potentiometer on the RF generator the overall resolution linearity can be adjusted.<\/p>\n<p><strong>11.<\/strong> Set up two survey windows in AugerScan. One will be set from 90 to 100 AMU and the other from 185 to 195 AMU. Initially set the resolution in the software to the last known good value or the value that is written on the front of the RF generator. Use an AMU of .05 and a time per step of 50 ms, one sweep.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?ssl=1\"><img data-attachment-id=\"1849\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/10-percent-moly-peaks\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=538%2C283&amp;ssl=1\" data-orig-size=\"538,283\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"10 percent moly peaks\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=300%2C157&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=538%2C283&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1849 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=538%2C283&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg\" alt=\"10 percent moly peaks\" width=\"538\" height=\"283\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?w=538&amp;ssl=1 538w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?resize=300%2C157&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?resize=500%2C263&amp;ssl=1 500w\" sizes=\"(max-width: 538px) 100vw, 538px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1849\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/10-percent-moly-peaks\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=538%2C283&amp;ssl=1\" data-orig-size=\"538,283\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"10 percent moly peaks\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=300%2C157&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?fit=538%2C283&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1849 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?resize=538%2C283&#038;ssl=1\" alt=\"10 percent moly peaks\" width=\"538\" height=\"283\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?w=538&amp;ssl=1 538w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?resize=300%2C157&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/10-percent-moly-peaks.jpg?resize=500%2C263&amp;ssl=1 500w\" sizes=\"(max-width: 538px) 100vw, 538px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p><strong>12.<\/strong> Acquire both surveys (one at a time).<\/p>\n<p>In each window the goal is to have approximately 10% peak to valley ratio (in the log scale). Adjust the resolution number for the 90 to 100 AMU peaks and the resolution fine for the 185 to 195 peaks. CW = more resolution. Remember that the Resolution number in the software will affect both the low and high peaks, and that the resolution fine pot on the RF generator will have more effect on the high peaks than the low peaks. Here is a table that might be helpful-<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?ssl=1\"><img data-attachment-id=\"1846\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/table\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=712%2C172&amp;ssl=1\" data-orig-size=\"712,172\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"table\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=300%2C72&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=584%2C141&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1846\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C141&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg\" alt=\"sims_adjustment_table\" width=\"584\" height=\"141\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?w=712&amp;ssl=1 712w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?resize=300%2C72&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?resize=500%2C120&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1846\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/table\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=712%2C172&amp;ssl=1\" data-orig-size=\"712,172\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"table\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=300%2C72&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?fit=584%2C141&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1846\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?resize=584%2C141&#038;ssl=1\" alt=\"sims_adjustment_table\" width=\"584\" height=\"141\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?w=712&amp;ssl=1 712w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?resize=300%2C72&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/table.jpg?resize=500%2C120&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p><strong><em>\u00a0Note that any time you change the resolution number for the 90 to 100 survey you need to change it to the same value for the 185 to 195 survey.<\/em><\/strong><\/p>\n<p>After a few iterations you will have the resolution set so that both the low and high AMUs are approximately 10% resolution.<\/p>\n<p><strong>13.<\/strong> Once the resolution linearity adjustment is complete, recheck the 16 AMU and 98 AMU peak positions and adjust if necessary.<\/p>\n<p>If you run out of range on the resolution fine adjustment or the cal low peak position contact RBD Instruments for additional procedures on how to correct those issues.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Calibrating the SIMSII RF Generator This post shows you how to calibrate the Balzers QMH RF generators that are used on the PHI SIMSII and 6300\/6600 series SIMS systems. 1. Load in a Molybdenum sample. Note that the sample mount &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/calibrating-the-simsii-rf-generator\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":1843,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[164],"tags":[202,232],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/12\/SIMS-II-adjustment-pots.jpg?fit=717%2C755&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-tD","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":4064,"url":"https:\/\/www.rbdinstruments.com\/blog\/a-numbers-for-phi-units\/","url_meta":{"origin":1837,"position":0},"title":"&#8220;A&#8221; Numbers for PHI units","author":"Randy","date":"September 9, 2024","format":false,"excerpt":"A numbers for older PHI surface analysis systems control units.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2024\/09\/32-150-A116.jpg?fit=1200%2C540&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1750,"url":"https:\/\/www.rbdinstruments.com\/blog\/high-voltage-connector-types\/","url_meta":{"origin":1837,"position":1},"title":"High voltage connector types","author":"Randy","date":"May 8, 2016","format":false,"excerpt":"There are primarily five different high voltage connectors that are used on surface analysis systems. This post will explain the differences between them and provide information on how to connect them. The high voltage connector types are: BNC MHV SHV 10kV SHV 20kV RF This connector is not really considered\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"SHV_connector","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/05\/SHV_connector.jpg?fit=1200%2C875&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/05\/SHV_connector.jpg?fit=1200%2C875&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/05\/SHV_connector.jpg?fit=1200%2C875&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/05\/SHV_connector.jpg?fit=1200%2C875&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":778,"url":"https:\/\/www.rbdinstruments.com\/blog\/augerscan-aes-xps-software-tutorial\/","url_meta":{"origin":1837,"position":2},"title":"AugerScan AES XPS software Tutorial","author":"Randy","date":"March 18, 2025","format":false,"excerpt":"AugerScan tutorial that now includes a video on AugerScan and AES","rel":"","context":"In &quot;Software for Surface Analysis Systems&quot;","block_context":{"text":"Software for Surface Analysis Systems","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/software-for-surface-analysis\/"},"img":{"alt_text":"Augerscan-Icons-vert","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/09\/Augerscan-Icons-vert.jpg?resize=350%2C200&ssl=1","width":350,"height":200},"classes":[]},{"id":1100,"url":"https:\/\/www.rbdinstruments.com\/blog\/calibrating-9103-usb-picoammeter\/","url_meta":{"origin":1837,"position":3},"title":"Calibrating the 9103 USB Picoammeter","author":"Joe Caterinicchio","date":"June 3, 2014","format":false,"excerpt":"Calibrating the 9103 USB Picoammeter The calibration procedure provided with the 9103 USB Picoammeter is written for use with a Keithley 220 current source and is set up for semi-automatic operation. But what if you want to use another brand or type of current source? The following procedure shows you\u2026","rel":"","context":"In &quot;9103 USB Picoammeter&quot;","block_context":{"text":"9103 USB Picoammeter","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/9103-usb-picoammeter\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/06\/9103-calibration-image.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/06\/9103-calibration-image.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/06\/9103-calibration-image.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/06\/9103-calibration-image.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":1184,"url":"https:\/\/www.rbdinstruments.com\/blog\/xps-aes-peak-linearity-adjustments\/","url_meta":{"origin":1837,"position":4},"title":"XPS and AES peak linearity adjustments","author":"Randy","date":"July 21, 2014","format":false,"excerpt":"This post is a compilation of some calibration tech tips that I have written over the years. The procedures listed below explain how to calibrate the following systems and units: 5600 and 5400 XPS systems, Double pass CMA XPS analyzers Scanning Auger system, Auger analyzers WARNING: Some of these procedures\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"xps-copper-gold-peaks","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/xps-copper-gold-peaks.jpg?fit=717%2C471&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/xps-copper-gold-peaks.jpg?fit=717%2C471&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/xps-copper-gold-peaks.jpg?fit=717%2C471&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1986,"url":"https:\/\/www.rbdinstruments.com\/blog\/loading-phi-specimen-mounts\/","url_meta":{"origin":1837,"position":5},"title":"Loading PHI specimen mounts","author":"Randy","date":"June 6, 2017","format":false,"excerpt":"Loading PHI specimen mounts This post explains the mechanics involved with loading PHI specimen mounts used on most of the older to current Physical Electronics surface analysis systems. There are a number of different sizes and shapes of specimen (sample) mounts and they all use the same basic mounting and\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"one_inch_sample_puck_above_specimen_stage","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/06\/one_inch_sample_puck_above_specimen_stage.jpg?fit=486%2C324&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1837"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=1837"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1837\/revisions"}],"predecessor-version":[{"id":1853,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1837\/revisions\/1853"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/1843"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=1837"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=1837"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=1837"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}