{"id":1779,"date":"2016-06-02T16:18:19","date_gmt":"2016-06-02T23:18:19","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=1779"},"modified":"2016-06-02T16:18:19","modified_gmt":"2016-06-02T23:18:19","slug":"72-100-voltage","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/","title":{"rendered":"72-100 voltage"},"content":{"rendered":"<p>I recently encountered an unusual problem on two different 660 scanning auger systems within a period of a few months where the 72-100 multiplier voltage would drop off after a period of time.<\/p>\n<p>On one system the 97 SED preamp would fail after a few minutes of warm up time, resulting in no image. On the other system the 96A V\/F preamplifier would fail after 20 to 30 minutes, resulting in no AES data.<\/p>\n<p>In both cases I measured the multiplier supply voltages at the preamp and the voltages were there when functioning and not there or unstable when in the failed mode. So easy enough, that should be a problem on the 72-100 board. To verify that the problem was on the board I swapped the 72-100 board with the other one. For those of you not familiar with the 660 scanning auger system or the 72-100 electron multiplier supply, there are two 72-100s on a 660. One is for the CMA AES analyzer and the other one is for the 97 SED preamplifier. The boards are identical and can be swapped out by physically moving the boards and changing an address switch.<\/p>\n<p>Swapping the 72-100 boards produced the same result. That meant that the problem was not on the board, so the next most likely thing was the SHV multiplier supply cable. Those cables checked out fine with an ohmmeter but to be sure I also swapped the cables out and the problem was still there.<\/p>\n<p>So as unlikely as it was, the only thing left was the 72-100 mother board.\u00a0\u00a0 I pulled out the mother board looking for a bent pin or something that could explain the problem and did in fact find that one of the 1K ohm filter resistors was out of spec and looked like it had been running hot. Moving it slightly caused the resistor to break in two.<\/p>\n<p>Here is a picture of the motherboard that shows the resistor:<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?ssl=1\"><img data-attachment-id=\"1777\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/72-100-r1-r2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=774%2C1032&amp;ssl=1\" data-orig-size=\"774,1032\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.2&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SCH-I545&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1452783577&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.2&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.0666666666667&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"72-100 R1 R2\" data-image-description=\"&lt;p&gt;72-100 R1 R2&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;72-100 R1 R2&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=225%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=584%2C779&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-large wp-image-1777\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C778&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2-768x1024.jpg\" alt=\"72-100 R1 R2\" width=\"584\" height=\"778\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?resize=768%2C1024&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?resize=225%2C300&amp;ssl=1 225w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?w=774&amp;ssl=1 774w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1777\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/72-100-r1-r2\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=774%2C1032&amp;ssl=1\" data-orig-size=\"774,1032\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;2.2&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;SCH-I545&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;1452783577&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;4.2&quot;,&quot;iso&quot;:&quot;200&quot;,&quot;shutter_speed&quot;:&quot;0.0666666666667&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"72-100 R1 R2\" data-image-description=\"&lt;p&gt;72-100 R1 R2&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;72-100 R1 R2&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=225%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?fit=584%2C779&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-large wp-image-1777\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?resize=584%2C778&#038;ssl=1\" alt=\"72-100 R1 R2\" width=\"584\" height=\"778\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?resize=768%2C1024&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?resize=225%2C300&amp;ssl=1 225w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-R1-R2.jpg?w=774&amp;ssl=1 774w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>72-100 R1 R2<\/p>\n<p>And here is the schematic that shows both resistors:<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?ssl=1\"><img data-attachment-id=\"1778\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/untitled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=766%2C699&amp;ssl=1\" data-orig-size=\"766,699\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"72-100 mother board schematic\" data-image-description=\"&lt;p&gt;72-100 mother board schematic&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;72-100 mother board schematic&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=300%2C273&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=584%2C533&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-full wp-image-1778\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C533&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg\" alt=\"72-100 mother board schematic\" width=\"584\" height=\"533\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?w=766&amp;ssl=1 766w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?resize=300%2C273&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?resize=328%2C300&amp;ssl=1 328w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1778\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/untitled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=766%2C699&amp;ssl=1\" data-orig-size=\"766,699\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"72-100 mother board schematic\" data-image-description=\"&lt;p&gt;72-100 mother board schematic&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;72-100 mother board schematic&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=300%2C273&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?fit=584%2C533&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-full wp-image-1778\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?resize=584%2C533&#038;ssl=1\" alt=\"72-100 mother board schematic\" width=\"584\" height=\"533\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?w=766&amp;ssl=1 766w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?resize=300%2C273&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/Untitled.jpg?resize=328%2C300&amp;ssl=1 328w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>72-100 mother board schematic<\/p>\n<p>Replacing the defective resistor solved the problem. What was interesting for me with this problem was that in over 37 years of working with PHI surface analysis systems I had never seen this problem before, and then I saw the same problem on two different systems within a short period of time.<\/p>\n<p>If you have an intermittent SED image or AES data problem on your 660 (or 4000 series) scanning auger system, keep this solution in mind as something to check.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>I recently encountered an unusual problem on two different 660 scanning auger systems within a period of a few months where the 72-100 multiplier voltage would drop off after a period of time. On one system the 97 SED preamp &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/72-100-voltage\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":1776,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[162],"tags":[224,100,225],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2016\/06\/72-100-mother-board.jpg?fit=774%2C1032&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-sH","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":3199,"url":"https:\/\/www.rbdinstruments.com\/blog\/72-100-electron-multiplier-supply-notes\/","url_meta":{"origin":1779,"position":0},"title":"72-100 Electron Multiplier Supply Notes","author":"Randy","date":"April 2, 2021","format":false,"excerpt":"Notes on to test and repair the 72-100 electron multiplier supply used on older PHI Auger systems.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/Extended-72-100-board.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":860,"url":"https:\/\/www.rbdinstruments.com\/blog\/how-to-test-a-97-sed-preamplifier\/","url_meta":{"origin":1779,"position":1},"title":"How to test a 97 SED preamplifier","author":"Randy","date":"November 8, 2013","format":false,"excerpt":"Older PHI scanning auger systems use the model 97 SED preamplifier to obtain secondary electron images. Occasionally you will not be able to get a TV image on your scanning auger system but are not sure what the problem is. This post will explain how to test the 97 SED\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"97-sed-col-tab","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/11\/97-sed-col-tab.jpg?fit=401%2C266&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1519,"url":"https:\/\/www.rbdinstruments.com\/blog\/660-scanning-auger-bake-out-procedure\/","url_meta":{"origin":1779,"position":2},"title":"660 Scanning Auger Bake-out procedure","author":"Randy","date":"June 15, 2015","format":false,"excerpt":"660 Scanning Auger Bake-out procedure. This procedure is written for the PHI 660 scanning Auger system but also works for the 600 scanning Auger systems as well. Ion pumps must be on and the system vacuum should be in the mid to low 10-7 range (or better).. Make sure that\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":879,"url":"https:\/\/www.rbdinstruments.com\/blog\/32-100-electron-multiplier-supply-digital-mode\/","url_meta":{"origin":1779,"position":3},"title":"32-100 electron multiplier supply digital mode","author":"Randy","date":"November 26, 2013","format":false,"excerpt":"The model 32-100 electron multiplier supply is used on older Physical Electronics Auger electron spectroscopy and X-ray photoelectron spectroscopy surface analysis systems to control the electron multiplier voltage. When using the 32-100 electron multiplier supply in the digital mode (the software automatically sets the electron multiplier voltage)\u00a0 the auto-ems box\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"32100-switch-positions","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/11\/32100-switch-positions.jpg?fit=546%2C564&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1616,"url":"https:\/\/www.rbdinstruments.com\/blog\/32-100-repair-calibration\/","url_meta":{"origin":1779,"position":4},"title":"32-100 Repair &#038; Calibration","author":"Randy","date":"January 3, 2016","format":false,"excerpt":"32-100 Repair & Calibration The 32-100 electron multiplier supply is used on many older PHI Auger and XPS surface analysis systems. This post has information about the 32-100 repair and calibration. Command Structure Use the diagnostic dialog box in Augerscan software to load the digital values per the 32-100 calibration\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":3301,"url":"https:\/\/www.rbdinstruments.com\/blog\/testing-the-fuses-on-the-card-rack-power-supply-outputs\/","url_meta":{"origin":1779,"position":5},"title":"Testing the fuses on the Card Rack power supply outputs","author":"Randy","date":"May 10, 2021","format":false,"excerpt":"The fuse strip located near the hinge on the OEM supply door routes the +5, +15, and -15 voltages from the OEM supply to the card rack unit motherboards. Each card rack unit has a dedicated section of fuse strip and specific fuse values. When troubleshooting problems with card rack\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/05\/reference-points.jpg?fit=768%2C1024&ssl=1&resize=700%2C400 2x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1779"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=1779"}],"version-history":[{"count":1,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1779\/revisions"}],"predecessor-version":[{"id":1780,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1779\/revisions\/1780"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/1776"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=1779"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=1779"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=1779"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}