{"id":1562,"date":"2015-09-03T16:46:15","date_gmt":"2015-09-03T23:46:15","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=1562"},"modified":"2015-09-03T16:46:15","modified_gmt":"2015-09-03T23:46:15","slug":"kratos-monochromator-adjustment","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/","title":{"rendered":"Kratos Monochromator adjustment"},"content":{"rendered":"<p>This Kratos monochromator adjustment procedure is used on the Kratos Axis, 165, Ultra and Nova monochromators.<\/p>\n<p>Steps involved in optimizing the performance of the Al Monochromatic x-ray source:<\/p>\n<p>First, load a sample bar containing a grid alongside a clean Ag foil sample at the same height.<\/p>\n<ol>\n<li>Using the Mg source image the grid to locate the height of the auto stage that gives the best focus. Move across to the Ag foil sample.<\/li>\n<li>Using the Mg anode find the BE position of the Ag 3d 5\/2 peak. This method requires that the calibration of the energy scale of the instrument is correct.<\/li>\n<li>Using the BE position obtained with the Mg dual anode switch to the mono source and begin tuning the source position in conjunction with the crystal settings as detailed below.<\/li>\n<li>The image below shows the mono crystal adjustment mechanism. First undo the locking screw. (Rather than just loosening the locking screw, it sometimes helps to use the locking screw as a counter drive against the micrometers to enable a finer adjustment on the crystal mechanism). Micrometer A tilts the crystal left and right moving the mono spot in the x-axis on the auto stage. This can be carefully adjusted to optimize the count rate on its own (no interaction with other parameters). Similarly, there is a tilt adjustment on the mono source that changes the source to crystal distance that can be adjusted for maximum count rate at this point.<\/li>\n<li>The micrometer on top of the mechanism (marked \u201cDo not adjust\u201d in the image) changes the distance between the crystal and the sample. (After initial factory set up \u2026i.e. for normal maintenance procedures\u2026anode changes etc&#8230; this will not need re-tuning).<\/li>\n<li>Micrometer B tilts the crystal to move the mono spot along the y-axis of the auto stage (towards and away from the dual anode source). This micrometer always needs to be adjusted in conjunction with the linear movement of the mono x-ray source itself. When the mono source position is changed it moves the refracted beam on the sample so the tilt adjustment is required to move the mono spot back to the analysis position to maximize the count rate.<\/li>\n<li>The linear drive on the mono source itself should be adjusted in small increments. At each new position micrometer B is used to move the beam spot back to the analysis position (by maximizing the count rate). The best method is to tilt the crystal carefully through the maximum once or twice and make a note of the best count rate obtained.<\/li>\n<\/ol>\n<p>Next, move the mono source a short distance (this will decrease the count rate). Now repeat the adjustment on Micrometer B to bring the count rate to a new maximum. If the new maximum is higher than the previous one continue the process moving the source in the same direction until the new maximum is lower than the previous one. Then move the source back to the previous setting and re-optimize.<\/p>\n<p>If at the start of the procedure the second maximum is lower than the first, reverse the direction of movement of the mono source and continue the process in that direction to find the best \u201cmaximum\u201d count rate.<\/p>\n<p>Once the set up is close to optimum performance more improvement can sometimes be obtained by setting up the tuning using the 110um selected area instead of large area \u201chybrid\u201d mode. (Starting with the 110um aperture is sometimes difficult due to the lower counting rates).<\/p>\n<p>The locking screw should finally be re-tightened in conjunction with small final adjustments of the micrometers to maintain the maximum count rate. The locking screw may push the crystal position slightly as it is tightened.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?ssl=1\"><img data-attachment-id=\"1568\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-micrometers\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=800%2C600&amp;ssl=1\" data-orig-size=\"800,600\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-micrometers\" data-image-description=\"&lt;p&gt;monochromator-micrometers&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=300%2C225&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=584%2C438&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1568 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C438&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg\" alt=\"monochromator-micrometers\" width=\"584\" height=\"438\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?w=800&amp;ssl=1 800w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?resize=300%2C225&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?resize=400%2C300&amp;ssl=1 400w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1568\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-micrometers\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=800%2C600&amp;ssl=1\" data-orig-size=\"800,600\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-micrometers\" data-image-description=\"&lt;p&gt;monochromator-micrometers&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=300%2C225&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=584%2C438&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1568 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?resize=584%2C438&#038;ssl=1\" alt=\"monochromator-micrometers\" width=\"584\" height=\"438\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?w=800&amp;ssl=1 800w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?resize=300%2C225&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?resize=400%2C300&amp;ssl=1 400w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p><strong>Installing parts using Au seals.<\/strong><\/p>\n<p>When installing using Au wire seals it is essential to locate the seal in the correct position. The seal should be resting snugly on the shoulder formed above the main flange.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?ssl=1\"><img data-attachment-id=\"1565\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" data-orig-size=\"215,139\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket\" data-image-description=\"&lt;p&gt;monochromator-gasket&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft size-full wp-image-1565\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=215%2C139&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg\" alt=\"monochromator-gasket\" width=\"215\" height=\"139\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1565\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" data-orig-size=\"215,139\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket\" data-image-description=\"&lt;p&gt;monochromator-gasket&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?fit=215%2C139&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft size-full wp-image-1565\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket.jpg?resize=215%2C139&#038;ssl=1\" alt=\"monochromator-gasket\" width=\"215\" height=\"139\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>This way when the 2 surfaces come together the Au wire is \u2018boxed\u201d in and compressed.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?ssl=1\"><img data-attachment-id=\"1566\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket-compressed\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" data-orig-size=\"195,76\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket-compressed\" data-image-description=\"&lt;p&gt;monochromator-gasket-compressed&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft size-full wp-image-1566\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=195%2C76&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg\" alt=\"monochromator-gasket-compressed\" width=\"195\" height=\"76\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1566\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket-compressed\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" data-orig-size=\"195,76\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket-compressed\" data-image-description=\"&lt;p&gt;monochromator-gasket-compressed&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?fit=195%2C76&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft size-full wp-image-1566\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-compressed.jpg?resize=195%2C76&#038;ssl=1\" alt=\"monochromator-gasket-compressed\" width=\"195\" height=\"76\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>If the Au seal is stretched or is not a tight fit on the shoulder and therefore \u201changs\u201d over the edge of the shoulder above the flange it will be pinched as the flanges are pulled together and <strong>NO SEAL CAN BE ACHIEVED NO MATTER HOW MUCH YOU TIGHTEN THE BOLTS!<\/strong><\/p>\n<p>Since the seal cannot be seen after mating up the flanges the only indication of whether the Au seal is in place is the gap between the 2 flanges when they are pushed together. There should only be a very small gap between the flanges &lt;~1mm. If the gap is larger it means the Au seal is pinched.<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?ssl=1\"><img data-attachment-id=\"1567\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket-incorrect\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" data-orig-size=\"205,134\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket-incorrect\" data-image-description=\"&lt;p&gt;monochromator-gasket-incorrect&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft size-full wp-image-1567\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=205%2C134&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg\" alt=\"monochromator-gasket-incorrect\" width=\"205\" height=\"134\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1567\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/monochromator-gasket-incorrect\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" data-orig-size=\"205,134\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"monochromator-gasket-incorrect\" data-image-description=\"&lt;p&gt;monochromator-gasket-incorrect&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?fit=205%2C134&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft size-full wp-image-1567\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-gasket-incorrect.jpg?resize=205%2C134&#038;ssl=1\" alt=\"monochromator-gasket-incorrect\" width=\"205\" height=\"134\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<h2>Water Box Maintenance<\/h2>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?ssl=1\"><img data-attachment-id=\"1564\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/kratos-water-box\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=763%2C807&amp;ssl=1\" data-orig-size=\"763,807\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Kratos-water-box\" data-image-description=\"&lt;p&gt;Kratos-water-box&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=283%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=584%2C618&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1564 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C618&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg\" alt=\"Kratos-water-box\" width=\"584\" height=\"618\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?w=763&amp;ssl=1 763w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?resize=283%2C300&amp;ssl=1 283w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1564\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/kratos-water-box\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=763%2C807&amp;ssl=1\" data-orig-size=\"763,807\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Kratos-water-box\" data-image-description=\"&lt;p&gt;Kratos-water-box&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=283%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?fit=584%2C618&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1564 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?resize=584%2C618&#038;ssl=1\" alt=\"Kratos-water-box\" width=\"584\" height=\"618\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?w=763&amp;ssl=1 763w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-water-box.jpg?resize=283%2C300&amp;ssl=1 283w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>With 15kV applied at the anode the water lines act as a conductive path to ground. The first point of contact to ground is at the outlet pipe from the solenoid valve (above the valve in the water box. (The water tube goes from there to the supply pipe on the anode). The other point is where the return line from the anode attaches to the flow switch (below the flow switch)<\/p>\n<p><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?ssl=1\"><img data-attachment-id=\"1563\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/kratos-anode-corrosion-points\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=679%2C404&amp;ssl=1\" data-orig-size=\"679,404\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Kratos-anode-corrosion-points\" data-image-description=\"&lt;p&gt;Kratos-anode-corrosion-points&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=300%2C178&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=584%2C347&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy alignleft wp-image-1563 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C347&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg\" alt=\"Kratos-anode-corrosion-points\" width=\"584\" height=\"347\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?w=679&amp;ssl=1 679w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?resize=300%2C178&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?resize=500%2C297&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"1563\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/kratos-anode-corrosion-points\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=679%2C404&amp;ssl=1\" data-orig-size=\"679,404\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"Kratos-anode-corrosion-points\" data-image-description=\"&lt;p&gt;Kratos-anode-corrosion-points&lt;\/p&gt;\n\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=300%2C178&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?fit=584%2C347&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"alignleft wp-image-1563 size-full\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?resize=584%2C347&#038;ssl=1\" alt=\"Kratos-anode-corrosion-points\" width=\"584\" height=\"347\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?w=679&amp;ssl=1 679w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?resize=300%2C178&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/Kratos-anode-corrosion-points.jpg?resize=500%2C297&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>&nbsp;<\/p>\n<p>These are the most like places for corrosion due to the voltage difference. Check and clean out or replace these fittings to bring up the flow rate.<\/p>\n<p>Also remember to check the in-line filter. If there is a drop in pressure across the filter as indicated by the pressure gauges, the filter may need to be replaced.<\/p>\n<p><strong>\u00a0NEVER LOWER THE SLIDER ON THE FLOW SWITCH TO GET BACK INTO OPERATION IF THE X-RAY SOURCE HIGH VOLTAGE WON\u2019T COME ON. OPERATING THE SOURCES AT LOWER FLOW THAN THE FACTORY RATING WILL CAUSE SHORTENED ANODE LIFETIMES.<\/strong><\/p>\n","protected":false},"excerpt":{"rendered":"<p>This Kratos monochromator adjustment procedure is used on the Kratos Axis, 165, Ultra and Nova monochromators. Steps involved in optimizing the performance of the Al Monochromatic x-ray source: First, load a sample bar containing a grid alongside a clean Ag &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/kratos-monochromator-adjustment\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":1568,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[160],"tags":[198,201,200,199],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/09\/monochromator-micrometers.jpg?fit=800%2C600&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-pc","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":2992,"url":"https:\/\/www.rbdinstruments.com\/blog\/monochromator-upgrade-for-phi-5600-xps-system\/","url_meta":{"origin":1562,"position":0},"title":"Monochromator upgrade for PHI 5600 XPS system","author":"Randy","date":"September 9, 2020","format":false,"excerpt":"Installing a monochromator This blog post shows the steps required to install a monochromator upgrade on a PHI 5600 XPS system. The components of a monochromator upgrade include the monochromator housing, the crystal substrate and aligner, the mono X-ray source, and the X-ray source aligner.\u00a0\u00a0\u00a0 If your X-ray source is\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/09\/monochromator-schematic.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/09\/monochromator-schematic.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/09\/monochromator-schematic.jpg?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":2284,"url":"https:\/\/www.rbdinstruments.com\/blog\/10-610-monochromator-x-ray-source-anode-replacement-procedure\/","url_meta":{"origin":1562,"position":1},"title":"10-610 monochromator x-ray source anode replacement procedure","author":"Randy","date":"April 10, 2018","format":false,"excerpt":"This post will show how to replace the 10-610 monochromator X-ray source anode and filaments.\u00a0 The 10-610 monochromator X-ray source is used in conjunction with the 10-410 or 10-420 monochromator.\u00a0\u00a0 When replacing the anode you should also replace both filaments and the deionizer cartridge (located in the 16-0XX heat exchanger).\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"melted anode","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/04\/melted-anode.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/04\/melted-anode.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/04\/melted-anode.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":406,"url":"https:\/\/www.rbdinstruments.com\/blog\/bake-out-procedure-to-improve-base-vacuum\/","url_meta":{"origin":1562,"position":2},"title":"PHI 5600 XPS system Bakeout procedure to improve base vacuum","author":"Randy","date":"August 30, 2019","format":"gallery","excerpt":"Updated 8-30-19 This post explains the bakeout procedure to improve base vacuum in a PHI 5600 X-ray photoelectron spectrometer. The procedure is basically the same for most of the older PHI (Physical Electronics) Spherical Capacitive Analyzer XPS surface analysis systems. PHI X-ray photoelectron spectrometers generally have built-in radiant heaters as\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"bakeout blankets on chamber","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/04\/bakeout-blankets-on-chamber.jpg?fit=500%2C333&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1971,"url":"https:\/\/www.rbdinstruments.com\/blog\/x-ray-source-arcing\/","url_meta":{"origin":1562,"position":3},"title":"X-ray source arcing","author":"Randy","date":"May 8, 2017","format":false,"excerpt":"X-ray source arcing results in unstable XPS data and can also damage the x-ray source power supply or high voltage control. Typically X-ray source arcing is caused by contamination on the anode, a coated anode support ceramic (the football ceramic) or a loose filament. However one unseen cause of x-ray\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"burnt x-ray source connector","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/05\/burnt-x-ray-source-connetor.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":684,"url":"https:\/\/www.rbdinstruments.com\/blog\/xps-analyzer-focal-point\/","url_meta":{"origin":1562,"position":4},"title":"XPS analyzer focal point","author":"Randy","date":"July 16, 2013","format":false,"excerpt":"This post will explain how to find the analyzer focal point on a PHI 5000 series XPS system and then align the system microscope to that point. The general idea is that the lens on the SCA (spherical capacitive analyzer) has a very specific focal point where the highest counts,\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"slotted-silver-specimen","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2013\/07\/slotted-silver-specimen-mounted-on-recessed-sample-mount.jpg?fit=664%2C565&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":1629,"url":"https:\/\/www.rbdinstruments.com\/blog\/5400-ls-operating-procedure\/","url_meta":{"origin":1562,"position":5},"title":"5400 LS operating procedure","author":"Randy","date":"February 9, 2016","format":false,"excerpt":"The following 5400 LS operating procedures provide details of many aspects of how to operate a 5000 LS XPS system. PHI did not make very many of the 5400 LS (large sample) systems but the ones that are still in operation are real work horses. On a side note, I\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1562"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=1562"}],"version-history":[{"count":1,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1562\/revisions"}],"predecessor-version":[{"id":1569,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1562\/revisions\/1569"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/1568"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=1562"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=1562"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=1562"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}