{"id":1211,"date":"2014-09-11T06:03:15","date_gmt":"2014-09-11T13:03:15","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=1211"},"modified":"2021-04-28T19:28:38","modified_gmt":"2021-04-29T02:28:38","slug":"multi-channel-detector-continuity-tests","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/","title":{"rendered":"Multi channel detector continuity tests"},"content":{"rendered":"<p>The multi channel detector (MCD) assembly found on Physical Electronics 5500 through 5800 X-ray photoelectron spectroscopy (XPS) systems comprises two pieces \u2013 the channel plate assembly and the connection flange.<\/p>\n<p>If you follow the procedure in the 10-360 spherical capacitive analyzer manual you can replace the channel plates (Chevron type) without too much difficulty. Two things to remember are to use some small Allen wrenches to hold the pins up (that removes the spring tension) and to make sure that the little marks on the channel plates face each other. \u00a0This drawing might be helpful \u2013\u00a0<a href=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/09\/MCD-detail.pdf\">MCD detail<\/a><\/p>\n<p>But equally important as installing the channel plates correctly is ensuring that you have good continuity on all of the pins on the connection flange, and that you are making a good electrical contact with the channel plate assembly when you install the flange.<\/p>\n<p>Before you install the connection flange, short out the pins on the connector flange with aluminum foil as shown in the pictures below. Then use an ohmmeter to measure the continuity of each pin to the flange. Each pin should have a resistance of less than 2 ohms. Sometimes the pins can become oxidized and if so the resistance can be high, as high as Meg ohms. If the resistance on one or more pins is high, very carefully use a tweezers and pull up on the wire on the end that is close to the connector. Move it no more than 1\/16&#8243; up and down. If you move it up too high, the pin will come off the connector pin and it is very difficult to put those back on. Verify that each pin has good continuity and is all the way down before installing the connection flange.<\/p>\n<p>When you install the connection flange (after first installing the channel plate assembly inside the SCA) measure the capacitance of each pin to the chamber. They should all have the same capacitance, typically about 250 pf. The 4 middle pins are not used so their capacitance will be about 100 pf. If the other 16 pins do not all have the same capacitance, then the some of the spring contacts are not making contact with the channel plate assembly. Try rotating the flange a little bit. You may need to slide the connection flange back a little bit and then reseat it. Just be sure that each pin has the correct capacitance before tightening the connection flange.<\/p>\n<p>Finally, the pins on the connection flange may have been bent slightly during the continuity test. Visually inspect the pins and carefully slide on the MCD preamp to make sure that it will go on without bending any of the pins. Adjust the pins as needed until the MCD preamp can go on easily. The pins on the connection flange are very soft and can bend easily. It is possible to force the MCD preamp down on a bent pin and if so, the pin will usually break when you try to straighten it. So just be aware that this and you will not bend the MCD pins.<\/p>\n<p>\u00a0<\/p>\n<p>RBD Instruments provides the MCD channel plates and technical support to help you with the installation.<\/p>\n<p>\u00a0<\/p>\n<p class=\"jetpack-slideshow-noscript robots-nocontent\">This slideshow requires JavaScript.<\/p><div id=\"gallery-1211-1-slideshow\" class=\"jetpack-slideshow-window jetpack-slideshow jetpack-slideshow-black\" data-trans=\"fade\" data-autostart=\"1\" data-gallery=\"[{&quot;src&quot;:&quot;https:\\\/\\\/i0.wp.com\\\/www.rbdinstruments.com\\\/blog\\\/wp-content\\\/uploads\\\/2014\\\/09\\\/MCD-plates-oreintation1.jpg?fit=689%2C570\\u0026ssl=1&quot;,&quot;id&quot;:&quot;1219&quot;,&quot;title&quot;:&quot;MCD plates orientation&quot;,&quot;alt&quot;:&quot;MCD plates orientation&quot;,&quot;caption&quot;:&quot;MCD plates orientation&quot;,&quot;itemprop&quot;:&quot;image&quot;},{&quot;src&quot;:&quot;https:\\\/\\\/i0.wp.com\\\/www.rbdinstruments.com\\\/blog\\\/wp-content\\\/uploads\\\/2014\\\/09\\\/short-out-mcd-pins-with-aluminum-foil.jpg?fit=516%2C290\\u0026ssl=1&quot;,&quot;id&quot;:&quot;1217&quot;,&quot;title&quot;:&quot;short-out-mcd-pins-with-aluminum-foil&quot;,&quot;alt&quot;:&quot;short-out-mcd-pins-with-aluminum-foil&quot;,&quot;caption&quot;:&quot;short-out-mcd-pins-with-aluminum-foil&quot;,&quot;itemprop&quot;:&quot;image&quot;},{&quot;src&quot;:&quot;https:\\\/\\\/i0.wp.com\\\/www.rbdinstruments.com\\\/blog\\\/wp-content\\\/uploads\\\/2014\\\/09\\\/XPS-analyzer-tipped-back.jpg?fit=516%2C290\\u0026ssl=1&quot;,&quot;id&quot;:&quot;1218&quot;,&quot;title&quot;:&quot;XPS-analyzer-tipped-back&quot;,&quot;alt&quot;:&quot;&quot;,&quot;caption&quot;:&quot;SCA-tilted-back&quot;,&quot;itemprop&quot;:&quot;image&quot;},{&quot;src&quot;:&quot;https:\\\/\\\/i0.wp.com\\\/www.rbdinstruments.com\\\/blog\\\/wp-content\\\/uploads\\\/2014\\\/09\\\/resistance-should-be-one-ohm.jpg?fit=290%2C516\\u0026ssl=1&quot;,&quot;id&quot;:&quot;1216&quot;,&quot;title&quot;:&quot;resistance-should-be-one-ohm&quot;,&quot;alt&quot;:&quot;resistance-should-be-one-ohm&quot;,&quot;caption&quot;:&quot;resistance-should-be-one-ohm&quot;,&quot;itemprop&quot;:&quot;image&quot;},{&quot;src&quot;:&quot;https:\\\/\\\/i0.wp.com\\\/www.rbdinstruments.com\\\/blog\\\/wp-content\\\/uploads\\\/2014\\\/09\\\/measure-resistance-of-each-mcd-pin.jpg?fit=290%2C516\\u0026ssl=1&quot;,&quot;id&quot;:&quot;1215&quot;,&quot;title&quot;:&quot;measure-resistance-of-each-mcd-pin&quot;,&quot;alt&quot;:&quot;measure-resistance-of-each-mcd-pin&quot;,&quot;caption&quot;:&quot;measure-resistance-of-each-mcd-pin&quot;,&quot;itemprop&quot;:&quot;image&quot;}]\" itemscope itemtype=\"https:\/\/schema.org\/ImageGallery\"><\/div>\n<p>\u00a0<\/p>\n<p>\u00a0<\/p>\n<p>\u00a0<\/p>\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3270\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/mcd-feedthrough-pins-labled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"MCD-feedthrough-pins-labled\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3270\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3270\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/mcd-feedthrough-pins-labled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"MCD-feedthrough-pins-labled\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i1.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3270\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-feedthrough-pins-labled.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>MCD pins<\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image\"><img data-attachment-id=\"3271\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/mcd-flange-spring-contacts-labled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"MCD-flange-spring-contacts-labled\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"584\" height=\"584\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C584&#038;ssl=1\" data-original=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"lazy wp-image-3271\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><noscript><img data-attachment-id=\"3271\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/mcd-flange-spring-contacts-labled\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=1512%2C1512&amp;ssl=1\" data-orig-size=\"1512,1512\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;,&quot;orientation&quot;:&quot;0&quot;}\" data-image-title=\"MCD-flange-spring-contacts-labled\" data-image-description=\"\" data-image-caption=\"\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=300%2C300&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=584%2C584&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" width=\"1512\" height=\"1512\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?fit=584%2C584&amp;ssl=1\" alt=\"\" class=\"wp-image-3271\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?w=1512&amp;ssl=1 1512w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=150%2C150&amp;ssl=1 150w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=300%2C300&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=768%2C768&amp;ssl=1 768w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?resize=1024%2C1024&amp;ssl=1 1024w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/MCD-flange-spring-contacts-labled.jpg?w=1168&amp;ssl=1 1168w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/noscript><figcaption>MCD spring contact pins<\/figcaption><\/figure>\n","protected":false},"excerpt":{"rendered":"<p>The multi channel detector (MCD) assembly found on Physical Electronics 5500 through 5800 X-ray photoelectron spectroscopy (XPS) systems comprises two pieces \u2013 the channel plate assembly and the connection flange. If you follow the procedure in the 10-360 spherical capacitive &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/multi-channel-detector-continuity-tests\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":0,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[164],"tags":[154,153],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-jx","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":3234,"url":"https:\/\/www.rbdinstruments.com\/blog\/5600-xps-system-mcd-channel-plate-replacement-procedure\/","url_meta":{"origin":1211,"position":0},"title":"5600 XPS system MCD channel plate replacement procedure.","author":"Randy","date":"April 28, 2021","format":false,"excerpt":"Procedure to install channel plates in a PHI 5600, 5700 or 5800 XPS system","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/04\/remove-bottom-MCD-plate.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":4154,"url":"https:\/\/www.rbdinstruments.com\/blog\/contact-issue-with-mcd-ceramic-on-phi-5600-xps-system\/","url_meta":{"origin":1211,"position":1},"title":"Contact issue with MCD ceramic on PHI 5600 XPS system","author":"Randy","date":"January 21, 2025","format":false,"excerpt":"This blog post shows how to repair an open POS connector on a 16 channel MCD anode ceramic used in the MCD detector on PHI 5600 thru 5800 XPS systems.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1200%2C1004&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1200%2C1004&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1200%2C1004&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2025\/01\/Missing-gold-on-contact.jpg?fit=1200%2C1004&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":3760,"url":"https:\/\/www.rbdinstruments.com\/blog\/electron-multipliers-for-older-phi-aes-and-xps-systems\/","url_meta":{"origin":1211,"position":2},"title":"Electron multipliers for older PHI AES and XPS systems","author":"Randy","date":"June 15, 2023","format":false,"excerpt":"Part numbers for electron multipliers for older Physical Electronics Auger and X-ray photoelectron analyzers.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2023\/06\/image.png?fit=775%2C507&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":3449,"url":"https:\/\/www.rbdinstruments.com\/blog\/sca-charging-lens-elements\/","url_meta":{"origin":1211,"position":3},"title":"SCA charging lens elements","author":"Randy","date":"December 1, 2021","format":false,"excerpt":"This blog will describe the symptoms and solution for lens elements that charge up due to an oxidized graphite coating. Overview - The Physical Electronics SCA (Spherical Capaitive Analyzer) has a series of 2 or 4 lenses that focus electrons into the energy analyzer section of the SCA. The first\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2021\/11\/lens-close-up.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1393,"url":"https:\/\/www.rbdinstruments.com\/blog\/amplifier-discriminator-tests\/","url_meta":{"origin":1211,"position":4},"title":"Amplifier Discriminator Tests","author":"Randy","date":"March 9, 2015","format":false,"excerpt":"One of the troubleshooting techniques that can be used to test the counting circuitry of an auger or XPS system is called \u201ctickling the brick\u201d.\u00a0\u00a0 The term came about from when the early PHI systems used the Princeton Applied Research (PAR) 1120 amplifier discriminator for pulse count detection of electrons.\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"amplifier-discriminator","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/03\/1120-amplifier-discriminator.jpg?fit=536%2C263&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":2237,"url":"https:\/\/www.rbdinstruments.com\/blog\/how-an-electron-multiplier-works\/","url_meta":{"origin":1211,"position":5},"title":"How an electron multiplier works","author":"Randy","date":"February 6, 2018","format":false,"excerpt":"This post will explain the basic concept of how an electron multiplier works. Electron multipliers are used in surface analysis instruments to boost the detected signal to a level where it can be amplified and processed into data. For Auger Electron spectrometers and X-ray photo electron analyzers the detected signal\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"continuous dynode electron multiplier gain","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/02\/continuous-dynode-electron-multiplier-gain.jpg?fit=719%2C435&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/02\/continuous-dynode-electron-multiplier-gain.jpg?fit=719%2C435&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2018\/02\/continuous-dynode-electron-multiplier-gain.jpg?fit=719%2C435&ssl=1&resize=700%2C400 2x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1211"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=1211"}],"version-history":[{"count":2,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1211\/revisions"}],"predecessor-version":[{"id":3272,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1211\/revisions\/3272"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=1211"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=1211"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=1211"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}