{"id":1173,"date":"2014-07-04T10:38:06","date_gmt":"2014-07-04T17:38:06","guid":{"rendered":"http:\/\/www.rbdinstruments.com\/blog\/?p=1173"},"modified":"2014-11-06T09:48:05","modified_gmt":"2014-11-06T17:48:05","slug":"auto-valve-control-adjustment-procedure","status":"publish","type":"post","link":"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/","title":{"rendered":"Auto Valve Control adjustment procedure"},"content":{"rendered":"<p>The Auto Valve Control (AVC) controls all of the air valves on many PHI systems. It has built in logic that uses two thermocouple sensors and a mechanical switch to determine the status of the system.<\/p>\n<p><strong>There are four common problems with the AVC: <\/strong><\/p>\n<p>1. The V4 ion gun valve will not open in the automatic mode.<\/p>\n<p>2. The gate valve will not close when the probe is retracted.<\/p>\n<p>3. The V1 led on the remote is neither Red (closed) nor Green (open).<\/p>\n<p>4. One or more of the solenoids stick, or leak.<\/p>\n<p><strong>Problems\/Solutions: <\/strong><\/p>\n<p><strong>1. The V4 ion gun valve will not open in the Automatic mode. <\/strong><\/p>\n<p>There are two thermocouple gauge tubes that sense the vacuum of the turbo(s), but only one TC gauge. TC1 is located on top of the turbo and determines whether or not the turbo pump is up to full speed. TC2 is under the tabletop and measures the vacuum in the intro chamber. If the intro pumps down to 5 bars but the V4 valve will not open, either the TC gauge is out of adjustment, or TC1 is defective.<\/p>\n<p>To adjust the TC gauge:<\/p>\n<p>1. Remove the cover on the AVC and slide it out enough to work is able to access inside of the unit. On some systems, such as the 600 Multiprobe, you can remove the tabletops and come from the top down. On other systems, such as 5000 series XPS systems, you will need to slide the AVC out of the rack a little bit.<\/p>\n<p>2. Refer this adjustment to qualified personnel. With a DVM, measure the voltage on pins 1 and 2 of P14. This is the connector that comes from the TC gauge control output. Pump down the intro. After you have 5 bars, you should have 9 to 10 mV DC on pins 1 and 2 of P14. If not, there is a pot on the back of the TC gauge (hockey puck) that needs to be adjusted. There is a hole on the back of the AVC that might allow you to do the adjustment from the back of the AVC with a long, thin screwdriver. Usually this does not line up and you need to remove the 3 screws at the base (under side of the AVC) that hold the TC gauge to the AVC and rotate the TC gauge up to get to the pot. Adjust the pot for 9 to 10 mV. It goes from Zero mV (air) to 10 mV (vacuum). <strong>CAUTION! <\/strong>If you have to move the TC gauge exercise extreme caution! There is a 220-volt AC terminal strip located near the TC gauge. It is recommended that the AVC be shut down (close all valves and turn off all turbo pumps first) and un-plugged before tilting the TC gauge. Touching the metal case to the exposed 220-volt AC terminal strip will cause arcing, damage to the AVC and possible electrical shock!<\/p>\n<div id=\"attachment_1174\" style=\"width: 736px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?ssl=1\"><img aria-describedby=\"caption-attachment-1174\" data-attachment-id=\"1174\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/avc-control-board\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=726%2C351&amp;ssl=1\" data-orig-size=\"726,351\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"AVC-control-board\" data-image-description=\"&lt;p&gt;AVC-control-board&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;AVC-control-board&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=300%2C145&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=584%2C282&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-full wp-image-1174\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C282&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png\" alt=\"AVC-control-board\" width=\"584\" height=\"282\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?w=726&amp;ssl=1 726w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?resize=300%2C145&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?resize=500%2C241&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-1174\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"1174\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/avc-control-board\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=726%2C351&amp;ssl=1\" data-orig-size=\"726,351\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"AVC-control-board\" data-image-description=\"&lt;p&gt;AVC-control-board&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;AVC-control-board&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=300%2C145&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=584%2C282&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-full wp-image-1174\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?resize=584%2C282&#038;ssl=1\" alt=\"AVC-control-board\" width=\"584\" height=\"282\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?w=726&amp;ssl=1 726w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?resize=300%2C145&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?resize=500%2C241&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> AVC-control-board<\/p><\/div>\n<p>NOTE:<\/p>\n<p>RBD provides an upgrade to AVC units where the P14 and TG guage pots are mounted on the front panel of the AVC. We install this upgrade as part of a standard AVC repair.<\/p>\n<p>3. After you have the TC gauge adjusted for 10 mV, adjust R103 (The only adjustment pot in the AVC) for 4 bars on the AVC remote. The fifth bar is timed and comes up automatically after two minutes, provided that the vacuum stays at 4 bars. Find the threshold for 4 bars, and turn it an additional 1 full turn. There is some hysteresis in the adjustment.<\/p>\n<p>Calibration of the TC gauge is now complete. If the V4 valve still does not open in the automatic mode, then the Thermocouple tube TC1 located on the turbo is most likely defective or out of range. Since there is only one controller, both of the thermocouple tubes need to be matched somewhat closely in performance.<\/p>\n<p>You can purchase these tubes from RBD, our part number is DST06MRE.<\/p>\n<p><strong>2. The gate valve will not close when the probe is retracted. <\/strong><\/p>\n<p>If the V1 gate valve does not close automatically after retracting the intro probe, most likely the intro probe switch is broken.<\/p>\n<p>1. Remove the probe bnc cable from the intro (located under the intro chamber near the pumping line).<\/p>\n<p>2. Short the center pin on the cable to the outer shield on the cable.<\/p>\n<p>3. V1 should now close. If so, then the intro probe switch is broken and needs to be repaired\/replaced.<\/p>\n<p>You can operate the intro by shorting the cable to close the V1 valve until it is convenient to remove the intro from the gate valve and repair\/replace the switch (make sure the turbo pump is OFF before you remove the intro). PHI\u2019s part number for the switch is 613174. You can also usually repair or replace these with a piece of spring steel for much less than the price of a new switch.<\/p>\n<p>If you have a magnetic load lock arm then sometimes the magnetic sensing switch can fail. If so, contact RBD for a replacement.<\/p>\n<p><strong>3. The V1 led on the remote is not Red (closed) or Green (open). <\/strong><\/p>\n<p>Sometimes when there is a power outage the AVC will not be re-set properly unless the probe is fully retracted first.<\/p>\n<p>1. Make sure that all valves are closed.<\/p>\n<p>2. Pull the probe all the way out.<\/p>\n<p>3. Make sure that the AVC is in the Automatic (not manual) mode.<\/p>\n<p>4. Turn the AVC off.<\/p>\n<p>5. After one to two seconds, turn the AVC back on. Do not leave it off too long or the turbo pump will vent.<\/p>\n<p>The V1 light should now be red. If not, there is a problem with the AVC. Contact RBD Instruments for assistance at (541)330-0723 or at rbdinstruments dot com<\/p>\n<p><strong>4. Any one of the valves sticks, does not open, or leaks. <\/strong><\/p>\n<p>After some number of years, the solenoids that drive the air-actuated valves become dried up. The result is intermittent operation. Sometimes the valves will stick open, other times closed. You can take the solenoids apart and lubricate the seals with vacuum grease. This is somewhat difficult to do and can result in the valve being totally inoperable. The best solution is to replace the defective solenoid with a new one. RBD has these parts in stock at all times.<\/p>\n<p>The RBD part numbers are:<\/p>\n<p>V1 Solenoid \u00a0\u00a0\u00a0 T062-4E2RE<\/p>\n<p>V2 Solenoid \u00a0\u00a0\u00a0 T062E1-3-10-35RE<\/p>\n<p>V3,4,5,6\u00a0 Solenoid \u00a0\u00a0\u00a0 T062-4E1RE<\/p>\n<div id=\"attachment_1175\" style=\"width: 877px\" class=\"wp-caption alignleft\"><a href=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?ssl=1\"><img aria-describedby=\"caption-attachment-1175\" data-attachment-id=\"1175\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/solenoid-part-numbers\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=867%2C483&amp;ssl=1\" data-orig-size=\"867,483\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"solenoid-part-numbers\" data-image-description=\"&lt;p&gt;avc-solenoid-part-numbers&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;avc-solenoid-part-numbers&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=300%2C167&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=584%2C325&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"lazy size-full wp-image-1175\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/plugins\/jquery-image-lazy-loading\/images\/grey.gif?resize=584%2C325&#038;ssl=1\" data-original=\"https:\/\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png\" alt=\"avc-solenoid-part-numbers\" width=\"584\" height=\"325\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?w=867&amp;ssl=1 867w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?resize=300%2C167&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?resize=500%2C278&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><p id=\"caption-attachment-1175\" class=\"wp-caption-text\"><noscript><img data-attachment-id=\"1175\" data-permalink=\"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/solenoid-part-numbers\/\" data-orig-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=867%2C483&amp;ssl=1\" data-orig-size=\"867,483\" data-comments-opened=\"1\" data-image-meta=\"{&quot;aperture&quot;:&quot;0&quot;,&quot;credit&quot;:&quot;&quot;,&quot;camera&quot;:&quot;&quot;,&quot;caption&quot;:&quot;&quot;,&quot;created_timestamp&quot;:&quot;0&quot;,&quot;copyright&quot;:&quot;&quot;,&quot;focal_length&quot;:&quot;0&quot;,&quot;iso&quot;:&quot;0&quot;,&quot;shutter_speed&quot;:&quot;0&quot;,&quot;title&quot;:&quot;&quot;}\" data-image-title=\"solenoid-part-numbers\" data-image-description=\"&lt;p&gt;avc-solenoid-part-numbers&lt;\/p&gt;\n\" data-image-caption=\"&lt;p&gt;avc-solenoid-part-numbers&lt;\/p&gt;\n\" data-medium-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=300%2C167&amp;ssl=1\" data-large-file=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?fit=584%2C325&amp;ssl=1\" decoding=\"async\" loading=\"lazy\" class=\"size-full wp-image-1175\" src=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?resize=584%2C325&#038;ssl=1\" alt=\"avc-solenoid-part-numbers\" width=\"584\" height=\"325\" srcset=\"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?w=867&amp;ssl=1 867w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?resize=300%2C167&amp;ssl=1 300w, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/solenoid-part-numbers.png?resize=500%2C278&amp;ssl=1 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" data-recalc-dims=\"1\" \/><\/noscript><\/a> avc-solenoid-part-numbers<\/p><\/div>\n<p>&nbsp;<\/p>\n","protected":false},"excerpt":{"rendered":"<p>The Auto Valve Control (AVC) controls all of the air valves on many PHI systems. It has built in logic that uses two thermocouple sensors and a mechanical switch to determine the status of the system. There are four common &hellip; <a href=\"https:\/\/www.rbdinstruments.com\/blog\/auto-valve-control-adjustment-procedure\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":1174,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","footnotes":"","jetpack_publicize_message":"","jetpack_is_tweetstorm":false,"jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false}}},"categories":[164],"tags":[],"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2014\/07\/AVC-control-board.png?fit=726%2C351&ssl=1","jetpack_shortlink":"https:\/\/wp.me\/p2DEXo-iV","jetpack_sharing_enabled":true,"jetpack_likes_enabled":true,"jetpack-related-posts":[{"id":1890,"url":"https:\/\/www.rbdinstruments.com\/blog\/v4-differential-valve-problem\/","url_meta":{"origin":1173,"position":0},"title":"V4 Differential Valve Problem","author":"Randy","date":"February 4, 2017","format":false,"excerpt":"V4 Differential Valve Problem This post describes how to fix the problem of the V4 differential valve not opening on an AVC when the DIFFY PUMP ION GUN button on the AVC remote box is depressed.\u00a0 Look at the photos at the bottom of the post for additional information. The\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"clean_the_contacts","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2017\/02\/clean_the_contacts.jpg?fit=504%2C378&ssl=1&resize=350%2C200","width":350,"height":200},"classes":[]},{"id":3501,"url":"https:\/\/www.rbdinstruments.com\/blog\/avc-auto-valve-controller-theory-and-operation\/","url_meta":{"origin":1173,"position":1},"title":"AVC Auto Valve Controller Theory and Operation","author":"Randy","date":"February 15, 2022","format":false,"excerpt":"Theory and operation of Auto Valve controller (AVC) used on older Physical Electronics surface analysis systems.","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2022\/02\/AVC-controller.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2022\/02\/AVC-controller.jpg?fit=1200%2C1200&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2022\/02\/AVC-controller.jpg?fit=1200%2C1200&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2022\/02\/AVC-controller.jpg?fit=1200%2C1200&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":2538,"url":"https:\/\/www.rbdinstruments.com\/blog\/avc-solenoid-replacement-procedure\/","url_meta":{"origin":1173,"position":2},"title":"AVC solenoid replacement procedure","author":"Randy","date":"June 9, 2019","format":false,"excerpt":"This procedure describes how to replace the solenoids in the Auto Valve Controller (AVC) used on Physical Electronics PHI surface analysis systems such as XPS Photo-electron and scanning Auger electron spectrometers. The AVC provides control of the pneumatic valves on the system by using 24V DC solenoids to route air\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"remove screw","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2019\/06\/remove-screw.jpg?fit=972%2C648&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2019\/06\/remove-screw.jpg?fit=972%2C648&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2019\/06\/remove-screw.jpg?fit=972%2C648&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1548,"url":"https:\/\/www.rbdinstruments.com\/blog\/up-to-air\/","url_meta":{"origin":1173,"position":3},"title":"660 System up to air procedure","author":"Randy","date":"August 5, 2015","format":false,"excerpt":"Overview This procedure will allow you to safely bring the 660 scanning auger system up to air for maintenance. \u00a0 Shut down AugerMap and turn off the 20-610 electron gun control (filament down slowly) Also turn off the 20-622 gun control. Turn off the card rack power If V4 is\u2026","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"phi-660-scanning-auger","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/08\/SAM660.jpg?fit=1200%2C900&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/08\/SAM660.jpg?fit=1200%2C900&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/08\/SAM660.jpg?fit=1200%2C900&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2015\/08\/SAM660.jpg?fit=1200%2C900&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":2891,"url":"https:\/\/www.rbdinstruments.com\/blog\/argon-bottle-installation-procedure\/","url_meta":{"origin":1173,"position":4},"title":"Argon Bottle Installation Procedure","author":"Randy","date":"June 18, 2020","format":false,"excerpt":"This procedure explains how to install and pump out a 25cc argon gas bottle for an 04-303 or 06-350 ion gun used on PHI XPS and AES systems.","rel":"","context":"In &quot;General Optics and Vacuum&quot;","block_context":{"text":"General Optics and Vacuum","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/general-optics-and-vacuum\/"},"img":{"alt_text":"25cc-argon-bottle","src":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/06\/25cc-argon-bottle.jpg?fit=756%2C756&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/06\/25cc-argon-bottle.jpg?fit=756%2C756&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.rbdinstruments.com\/blog\/wp-content\/uploads\/2020\/06\/25cc-argon-bottle.jpg?fit=756%2C756&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1629,"url":"https:\/\/www.rbdinstruments.com\/blog\/5400-ls-operating-procedure\/","url_meta":{"origin":1173,"position":5},"title":"5400 LS operating procedure","author":"Randy","date":"February 9, 2016","format":false,"excerpt":"The following 5400 LS operating procedures provide details of many aspects of how to operate a 5000 LS XPS system. PHI did not make very many of the 5400 LS (large sample) systems but the ones that are still in operation are real work horses. On a side note, I\u2026","rel":"","context":"In &quot;Operation and Calibration Procedures&quot;","block_context":{"text":"Operation and Calibration Procedures","link":"https:\/\/www.rbdinstruments.com\/blog\/category\/operation-and-calibration-procedures\/"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]}],"_links":{"self":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1173"}],"collection":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/comments?post=1173"}],"version-history":[{"count":1,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1173\/revisions"}],"predecessor-version":[{"id":1176,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/posts\/1173\/revisions\/1176"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media\/1174"}],"wp:attachment":[{"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/media?parent=1173"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/categories?post=1173"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.rbdinstruments.com\/blog\/wp-json\/wp\/v2\/tags?post=1173"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}